Abstract
Purpose. The purpose of this study was to measure intravascular hematocrit values continuously by using a fiberoptic probe based on near-infrared photometry.
Methods. We produced a catheter 1.5 mm in diameter that use a pair of plastic fibers. One of the fibers, the measuring fiber, was used to measure the optical density of blood, and the other, the reference fiber, was used to decrease the signal-to-noise ratio. We employed an 805-nm laser diode as the light source. Two photodiodes were used to measure the intensity of the light transmitted through the two fibers, and the output signals were amplified and sent to a personal computer through an analog-to-digital converter.
Results. The hematocrit values obtained by this fiberoptic continuous measurement agreed well with those obtained by microcentrifugation within physiological ranges.
Conclusions. This method is effective for monitoring the rapid changes in hematocrit.
Similar content being viewed by others
References
Tanaka Y, Morimoto T, Watari H, Miyazaki M (1976) Continuous monitoring of circulating blood hematocrit. Jpn J Physiol 26:345–353
Leypoldt JK, Cheung AK, Steuer RR, Harris DH, Conis JM (1995) Determination of circulating blood volume by continuously monitoring hematocrit during hemodialysis. J Am Soc Nephrol 6:214–219
de Vries, JPPM, Olthof CG, Visser V, Kouw P, Es AV, Donker AJM, de Vries PMJP (1992) Continuous measurement of blood volume during hemodialysis by an optical method. ASAIO 38:M181-M185
Mancini E, Antonio S, Spongano M, Paolini F, Rossi M, Zucchelli P (1993) Continuous on-line optical absorbance recording of blood volume change during hemodialysis. Artif Organs 17(8): 691–694
Iwasaki H, Yoshizaki K, Koyano H (1992) Fiber-optic transmission catheter for regional venous oxygen saturation and blood flow. J Appl Physiol 72:1616–1621
Schallenberg U, Stiller S, Mann H (1987) A new method of continuous hemoglobinometric measurement of blood volume during hemodialysis. Life Supp Syst 5:293–305
Steuer RR, Harris DH, Conis JM (1993) A new optical technique for monitoring hematocrit and circulating blood volume: its application in renal dialysis. Dial Transplant 22: 260–265
Steinke JM, Shepherd AP (1987) Reflectance measurements of hematocrit and oxyhemoglobin saturation. Am J Physiol 253:H147-H153
Twersky V (1964) On propagation in random media of discrete scatterers. Proceedings of Symposia in Applied Mathematics (Providence, RI American Mathematical Society) 16:84–116
Loewinger E, Gordon A, Weinreb A, Gross J (1964) Analysis of a micromethod for transmission oximetry of whole blood. J Appl Physiol 19:1179–1184
Anderson NM, Sekelj P (1967) Light-absorbing and scattering properties of nonhaemolysed blood. Phys Med Biol 12:173–184
Tanaka Y (1979) Whole body transvascular filtration coefficient and interstitial space capacitance. Jpn J Physiol 29:181–193
Author information
Authors and Affiliations
About this article
Cite this article
Fukushima, H. Continuous measurement of hematocrit using an intravascular catheter equipped with a fiberoptic transmission cell. J Anesth 12, 87–94 (1998). https://doi.org/10.1007/BF02480778
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02480778