Abstract
Blood and dialysate flow patterns in hollow-fiber dialyzers are complicated, and hence the flow patterns and mass transfer are difficult to analyze theoretically. Consequently, dialyzers are usually developed by a trial-and-error method. We attempt to design dialyzers by computer simulation analysis in this work. Blood-side and dialysate-side flows were modeled using the Hagen-Poiseuille equation and the Blake-Kozeny equation, respectively. These flow patterns were evaluated as pressure drop and velocity distribution. The mass transfer rate was evaluated as solute clearance. Computed values of the pressure drops and clearance for urea and vitamin B12 were found to agree closely with those obtained experimentally. We evaluated the influences of the inner diameter of hollow fibers, module geometry, and void fraction on the pressure drop and clearance, and computer-aided design was performed.
Similar content being viewed by others
References
Sakai K. Technical determination of optimal dimensions of hollow fibre membranes for clinical dialysis. Nephrol Dial Transplant 1989;4(Suppl):73–77
Takesawa S, Terasawa M, Sakagami M, Kobayashi T, Hidai H, Sakai K. Nondestructive evaluation by X-ray computed tomography of dialysate flow patterns in capillary dialyzers. ASAIO J 1988;34:794–799
Bird RB, Stewart WE, Lightfoot EN. Transport phenomena. New York: Wiley, 1960:196–200
McAdams WH. Heat transmission, 3rd edn. New York: McGraw-Hill, 1954
Fukuda M, Hosoya N, Kanamori T, Sakai K, Nisikido J, Watanabe T, Fusimi F. Determination of optical fiber density of conventional and higher performance dialyzer. Artif Organs Today 1992;2:205–214
Ohmura T, Tasuguchi T, Nishikido J, Yamamoto T, Fushimi F, Nishida O, Sakai K. A new method of determining the solute permeability of hollow-fiber dialysis membranes by means of laser lights traveling along optic fibers. ASAIO J 1989;35:601–603
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Suzuki, Y., Kohori, F. & Sakai, K. Computer-aided design of hollow-fiber dialyzers. J Artif Organs 4, 326–330 (2001). https://doi.org/10.1007/BF02480026
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02480026