The effect of airway narrowing and dead space on the shape of the capnogram

  • Hiroyuki Kinoshita
  • Kenichi Sakamoto
  • Masaharu Ito
Original Articles
  • 41 Downloads

Abstract

We have investigated the effects of incomplete obstruction of the endotracheal tube and the amount of additional dead space between the endotracheal tube and the capnographic sampling adapter on the shape of the capnogram. A 9.0-mm endotracheal tube was connected to a 3-L reservoir bag filled from the bottom with 5% carbon dioxide and 95% oxygen. The narrowed adapter (internal diameter: 3.0, 4.0, 6.5, and 9.0 mm), the capnographic sampling adapter, and a semiclosed respiratory system were successively connected to this endotracheal tube. Additional dead space (0, 30, 62, 92, 124, or 154 ml) was inserted between the narrowed adapter and the capnographic sampling adapter. The reservoir bag was ventilated with the anesthesia ventilator (fresh gas flow, 6 L·min−1; tidal volume, 500 ml; respiratory rate, 10 min−1, and inspiratory-expiratory ratio, 1∶2). The capnogram from each initial ventilation was recorded and the peak carbon dioxide tension (Ppeak CO2) was also measured. The T90% value was defined as the time it took for the capnograph output to respond from 0% to 90% of the Ppeak CO2. The T90% value seen in a 3.0-mm adapter did not change compared with the value in a 9.0-mm adapter, when no additional dead space was connected between the endotracheal tube and the capnographic sampling adapter. Further, the slanting upstroke of the capnogram occurred only when the endotracheal tube narrowing and a large amount of dead space between the endotracheal tube and the capnographic sampling adapter coexisted. Thus, it is unlikely that incomplete obstruction of the endotracheal tube can easily be detected by the slanting upstroke of the capnogram.

Key words

Airway Capnography Capnogram Carbon dioxide tension Dead space 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Gravenstein JS, Paulus DA, Hayes TJ (1989) Capnography in clinical practice. Butterworths, Boston, pp 43–49Google Scholar
  2. 2.
    Good ML (1993) Principles and practice of capnography. Annual Refresher Course Lectures 431Google Scholar
  3. 3.
    Murray IP, Modell JH (1983) Early detection of endotracheal tube accidents by monitoring carbon dioxide concentration in respiratory gas. Anesthesiology 59:344–346, 1983PubMedGoogle Scholar
  4. 4.
    Badgwell JM, McLeod ME, Lerman J, Creighton RE (1987) Endtidal PCO2 measurements sampled at the distal and proximal ends of the endotracheal tube in infants and children. Anesth Analg 66:959–964PubMedCrossRefGoogle Scholar
  5. 5.
    Breen PH, Mazumdar B, Skinner SC (1994) Capnometer transport delay: measurement and clinical implications. Anesth Analg 78:584–586PubMedCrossRefGoogle Scholar
  6. 6.
    Brunner JX, Westenskow DR (1988) How the rise time of carbon dioxide analysers influences the accuracy of carbon dioxide measurements. Br J Anaesth 61:628–638PubMedGoogle Scholar

Copyright information

© JSA 1995

Authors and Affiliations

  • Hiroyuki Kinoshita
    • 1
  • Kenichi Sakamoto
    • 2
  • Masaharu Ito
    • 2
  1. 1.Department of AnesthesiaKurashiki Central HospitalKurashiki, OkayamaJapan
  2. 2.Department of AnesthesiaEhime Rousai HospitalNiihama, EhimeJapan

Personalised recommendations