Skip to main content
Log in

Dynamic calibration of manometer systems

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

Two simple and inexpensive methods for dynamic calibration of manometer systems for the clinical environment are described. One is a modification of a well known step-function technique, the other is a sinusoidal pressure generator using magnetic fluids. The practical use of both methods is described in detail, and a test of the accuracy of the methods is made. In particular, it is found that both methods are useful in the frequency range of at least 0 to 100 Hz, and that resonance frequencies could be determined with an error of about±5%, whereas the error in damping factor is greater, of the order of ±50%. The step-function technique is quicker and easier to apply under sterile conditions and is therefore recommended for wider use.

Sommaire

L'article décrit deux méthodes simples et non coûteuses de calibration dynamique des systèmes manométriques, applicables dans les milieux cliniques. La première est une modification d'une technique de fonction en échelon bien connue et la seconde est un générateur de pression sinusoïdale à partir de fluides magnétiques. L'usage pratique des deux méthodes est décrit en détail et des essais sur leur précisions sont effectués. On trouve en particulier que les deux méthodes sont pratiques dans la zone allant de 0 à 100 Hz au moins et que les fréquences de résonance peuvent être déterminées avec une erreur d'environ ±5% alors que l'erreur du facteur d'atténuation est plus élevée, de l'ordre de ±50%. La technique de la fonction en échelon est plus rapide et plus simple à utiliser sous milieu ambiant stérile, son usage est donc recommandé dans un plus grand nombre de cas.

Zusammenfassung

Zwei einfache, billige Verfahren zur dynamischen Eichung von Manometersystemen für den klinischen Gebrauch werden beschrieben. Das eine ist eine Abänderung eines wohlbekannten Stufenfunktionsverfahrens, das andere ist ein sinusähnlicher Druckgenerator mit Magnetflüssigkeiten. Die praktische Anwendung beider Verfahren wird genau beschrieben, und die Genauigkeit der Verfahren wird geprüft. Insbesondere wird festgestellt, daß beide Verfahren im Frequenzbereich von wenigstens 0 bis 100 Hz nütxlich sind und daß Resonanzfrequenzen mit einer Regelabweichung von etwa±5% bestimmt werden könnten, wogegen die Regelabweichung im Schwächungsfaktor größer ist, d.h. bei ±50% liegt. Das Stufenfunktionsverfahren ist schneller und leichter unter sterilen Bedingungen anzuwenden und wird daher für einen Gebrauch in größerem Umfang emfohlen.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Ball, G. andGabe, I. (1963) Sinusoidal pressure generator for testing differential manometers.Med. Electron. Biol. Eng. 1, 237–241.

    Article  Google Scholar 

  • Hansen, A. T. (1949) Pressure measurement in the human organism.Acta Physiol. Scand. 19, Suppl. 68, 7–230.

    Google Scholar 

  • Henry, W. L., Wilner, B. andHarrison, D. C. (1967) A calibrator for detecting bubbles in cardiac catheter-manometer systems.J. Appl. Physiol. 23, 1007–1009.

    Google Scholar 

  • Hök, B. (1975) New microtransducer for physiological pressure recordings.Med. and Biol. Eng. 13, 279–284.

    Article  Google Scholar 

  • Krovetz, L. J. andGoldbloom, S. D. (1974) Frequency content of intravascular and intracardiac pressures and their time derivatives.IEEE Trans. Biomed. Eng. BME-21, 498–501.

    Google Scholar 

  • McDonald, D. A. (1974)Blood flow in arteries, 2nd ed. Edward Arnold, London.

    Google Scholar 

  • Noble, F. W. (1959) A hydraulic pressure generator for testing the dynamic characteristics of blood pressure manometers.J. Lab. & Clin. Med. 54, 897–902.

    Google Scholar 

  • Shelton, C. D. andWatson, B. W. (1968) A pressure generator for testing the frequency response of catheter-transducer systems used for physiological pressure measurements.Phys. Med. Biol. 13, 523–528.

    Article  Google Scholar 

  • Stegall, H. F. (1967) A simple, inexpensive, sinusoidal pressure generator.J. Appl. Physiol. 22, 591–592.

    Google Scholar 

  • Vierhout, R. R. andVendrik, A. J. H. (1961) A hydraulic pressure generator for testing the dynamic characteristics of catheters and manometers.J. Lab. Clin. Med. 58, 330–333.

    Google Scholar 

  • Vierhout, R. R. andVendrik, A. J. H. (1964) On pressure generators for testing catheter manometer systems.Phys. Med. Biol. 10, 403–406.

    Article  Google Scholar 

  • Yanof, H. M., Rosen, A. L., McDonald, N. M. andMcDonald, D. A. (1963) A critical study of the response of manometers to forced oscillations.Phys. Med. Biol. 8, 407–422.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

The work was carried out at the Electronics Department, Institute of Technology, University of Uppsala

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hök, B. Dynamic calibration of manometer systems. Med. & biol. Engng. 14, 193–198 (1976). https://doi.org/10.1007/BF02478747

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02478747

Keywords

Navigation