Medical and biological engineering

, Volume 14, Issue 5, pp 565–569 | Cite as

Low-noise cable and slip-ring assembly for recording of small-amplitude physiological signals in chronic animals

  • A. T. Pryzbylik
  • R. C. Howe
Article

Abstract

A method is described for the construction of a multichannel low-noise cable and counter-balanced slip-ring device utilising commercially available components. This apparatus was used in conjunction with a system for recording microvolt physiological signals in unrestrained animals, and resulted in a substantial reduction of movement artefact. Although this method was used for recording the electroencephalogram (e.e.g.) in cats with chronically implanted electrodes, it is readily adaptable to other biomedical studies requiring artefact-free recording conditions.

Keywords

Artefact reduction Electroencephalogram Low-noise Slip-ring Swivel 

Sommaire

On décrit une méthode pour la fabrication d'un câble multicanaux à faible bruit et d'un dispositif à bague collectrice équilibrée en utilisant des composants disponsibles sur le marché. Cet appareil a été utilisé en conjugaison avec un système d'enregistrement des signaux physiologiques microvolts chez des animaux non restreints, et on a obtenu une réduction considérable d'artéfact de mouvement. Quoique cette méthode ait été utilisée pour l'enregistrement de lélectroencéphalogramme (EEG) des chats avec des électrodes implantées chroniquement, elle s'adapte facilement aux autres études biomédicales nécessitant des conditions d'enregistrement sans artéfact.

Zusammenfassung

Es wird win Verfahren zum Bau eines rauscharmen Mehrkanal-Kabels und eines abgeglichenen Schleifringelements unter Verwendung von im Handel verfügbaren Teilen beschrieben. Dieses Gerät wurde in Verbindung mit einem System zur Aufzeichnung von physiologischen Mikrovoltsignalen an unbehinderten Tieren werwendet und ergab eine wesentliche Senkung der Bewegungsartefakte. Obgleich dieses Verfahren zum Aufzeichnen von EEGs von Katzen mit chronisch eingepflanzten Elektroden verwendet wurde, läßt es sich ohne weiteres auf andere biologisch-medizinische Untersuchungen einstellen, für die artefaktfreie Aufnahmebedingungen erforderlich sind.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Ballintijn, C. M. (1965) Electric recording from unnarcotised, unrestrained small animals.Med. Electron. Biol. Eng.,3, 771.CrossRefGoogle Scholar
  2. Bloss, J. L., andPotts, W. J. (1974) A simple swivel for intracranial electrical stimulation.Physiol. Behav.13, 343.CrossRefGoogle Scholar
  3. Craggs, M. D. (1973) Miniature multichannel low-noise cable.Med. & Biol. Eng.11, 104.CrossRefGoogle Scholar
  4. Howard, B. R., andLornie, J. C. (1972) A minimalartefact lead for electroencephalography in the domestic fowl.Med. & Biol. Eng.10, 109.CrossRefGoogle Scholar
  5. Howe, R. C., andSterman, M. B. (1972) Corticalsubcortical EEG correlates of suppressed motor behaviour during sleep and waking in the cat.Electroenceph. Clin. Neurophysiol.32, 681–695.CrossRefGoogle Scholar
  6. Howe, R. C., andSterman, M. B. (1973) Somatosensory system evoked potentials during waking behaviour and sleep in the cat.Electroenceph. Clin. Neurophysiol.34, 605–618.CrossRefGoogle Scholar
  7. Kamp, A., Kok, M. L., andde Quartel, F. W. (1965) A multiwire cable for recording from moving subjects.Electroenceph. Clin. Neurophysiol.18, 422.CrossRefGoogle Scholar
  8. Mundl, W. J. (1967) An improved swivel for recording from behaving subjects.Electroenceph. Clin. Neurophysiol.23, 483.CrossRefGoogle Scholar
  9. Pryzbylik, A. T., andHowe, R. C. (1975) A system for biofeedback conditioning of electroencephalographic activity.Med. & Biol. Eng.13, 577–582.CrossRefGoogle Scholar
  10. Straw, R. N., McAdam, D., Berry, C. A., andMitchell, C. L. (1967) A simple cable for reduction of movement artefact in electroencephalographic recordings.Electroenceph. Clin. Neurophysiol.22, 90.CrossRefGoogle Scholar
  11. Waldron, K. andPhillips, A. G. (1972) A compact, maintenance free commutator for use in conjunction with chronic electrode preparations.Physiol. Behav.9, 881.CrossRefGoogle Scholar

Copyright information

© International Federation for Medical & Biological Engineering 1976

Authors and Affiliations

  • A. T. Pryzbylik
    • 1
  • R. C. Howe
    • 2
  1. 1.Department of Neurophysiology, Walter Reed Army Institute of ResearchWalter Reed Army Medical CenterWashington, D.C.
  2. 2.Department of Physiology and BioengineeringEastern Virgini Medical SchoolNorfolk