Medical and biological engineering

, Volume 13, Issue 4, pp 544–550 | Cite as

Radiotelemetry of avian shank temperature using a thin-film hybrid microcircuit

  • I. J. H. Duncan
  • J. H. Filshie
  • I. J. McGee
Article

Abstract

The shank temperature of a bird is one indicator of its behavioural state. A radiotelemetry system has been developed consisting of a thin-film radiotransmitter and a decoding circuit using pulse-ratio modulation of a subcarrier oscillator linearised for use with athermistor. The subcarrier frequency-modulates a 25–30 MHz carrier operating in near-field conditions. The transmitter can be used externally or may be implanted. Under experimental conditions temperature drift does not exceed 0·1 deg. C. Temperature changes in response to stress situations are reported and discussed.

Keywords

Behaviour Microelectronics Radiotelemetry Temperature 

Sommaire

La température du métatarse d'un oiseau est une indication de l'état de son comportement. On a développé un système radiotélémétrique consistant d'un radioémetteur en pellicule mince et d'un circuit décodeur utilisant une modulation de rapport de pulsation d'un oscillateur sous-porteur linéarisé pour usage avec thermistance. La fréquence du sous-porteur module un porteur de 25–30 MHz opérant sous des conditions de champ proche. L'émetteur peut être utilisé extérieurement ou peut être implanté. Dans les conditions expérimentales, la dérivation de température ne dépasse pas 0·4°C. On rapporte et on débat les changements de température résultant de contraintes.

Zusammenfassung

Die Unterschenkeltemperatur von Vögeln ist ein Zeichen für ihren Verhaltenszustand. Es wurde ein radiotelementrisches System entwickelt, das aus einem Dünnfilm-Funksender und einer Entschlüsselungsschaltung mit Impulsverhältnismodulation eines Zwischenträgeroszillators besteht, der zum Einsatz mit einem Thermistor linearisiert wird. Die Frequenz des Zwischenträgers moduliert einen Träger von 25–30 MHz, der unter Nahfeldbedingungen eingesetzt wird. Der Sender kann extern angewandt oder eingepflanzt werden. Unter experimentellen Bedingungen übersteigt die Temperaturabweichung nicht 0,1°C. Temperaturänderungen als Reaktion auf Überbeanspruchungen werden im Bericht erwähnt und besprochen.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Beneteau, P. J. andEvangelisti, A. (1963)An improved emitter-coupled multivibrator. SGS Fairchild Application Report AR-19.Google Scholar
  2. Beakley, W. R. (1951) The design of thermistor thermometers with linear calibration.J. Sci. Instrum. 28, 176–179.CrossRefGoogle Scholar
  3. Filshie, J. H. andMcGee, I. J. (1974a) Thin film hybrid circuits for biomedical radiotelemetry. Pt. 1-A film deposition process using electron-beam evaporation.Electron. Equip. News,15 (11), 92–94.Google Scholar
  4. Filshie, J. H. andMcGee, I. J. (1974b) Thin-film hybrid circuits for biomedical radiotelemetry. Pt. 2-Production of implantable devices.Electron. Equip. News 15, (12), 62–65.Google Scholar
  5. GPO (1968)Performance specification for radio equipment for medical and biological telemetry devices. HMSO SO Code No. 88-5438*.Google Scholar
  6. McKay, R. S. (1968)Biomedical telemetry. John Wiley, p. 53.Google Scholar
  7. Robinson, S. G. (1964) A temperature telemetry system with constant accuracy.Med. Electron. & Biol. Engng. 2, 81–83.CrossRefGoogle Scholar
  8. Voegli, F. andKraft, W. (1972) Multichannel telemetry of physiological parameters (body temperature, e.c.g., e.e.g.) in the rat. In:Biotelemetry. Eds. H. P.Kimmich andJ. A. Vos, 371–380. Meander NV.Google Scholar