Continuous mass spectrographic measurement of halothane partial pressure in blood

  • Melville Roberts
  • Ethan T. ColtonIII
  • Guy Owens
  • David D. Thomas
  • George M. Watkins
Article

Abstract

A method has been devised for the continuous measurement of halothane partial pressure in blood. The technique which was used in 10 dogs employs a silastic covered stainless-steel intra-arterial cannula through which blood gases pass into the analysing portion of a mass spectrometer where they are identified. Halothane produces a spectrographic peak at mass 31 which is proportional to its partial pressure. The peak at 31 results from the splitting of the halothane molecule into an ion composed of an atom of carbon and one of fluorine. The initial response time for the system is 0·4 min. The time from an initial changes to 63% of the final reading (1/e time) is 6·3 min. The calibration of the spectrometer remained stable for up to 7h, the longest test period.

Keywords

Blood halothane levels Mass spectrometer 

Sommaire

Une méthode a été mise au point pour mesurer en continu la pression partielle d'halothane dans le sang. La technique utilisée chez 10 chiens emploie une canule intra-artérielle en acier inoxydable et à revêtement silastic par l'intermédiaire de laquelle les gaz contenus dans le sang passent dans l'analyseur d'un spectromètre de masse où ils sont identifiés. L'halothane produit une crête spectrographique à la masse à 31 qui est proportionnelle à sa pression partielle. Cette crête à 31 a pour cause la division de la molécule d'halothane en un ion composé d'un atome de carbone et d'un atome de fluor. Le temps de réponse initial pour le système est de 0,4 min. La durée du passage d'une variation initiale demeure stable jusqu'à 7 heures, c'est-à-dire la période d'essai la plus longue.

Zusammenfassung

Es wurde ein Verfahren zur ständigen Messung des partiellen Drucks von Halothan im Blut gefunden. Das Verfahren, das an zehn Hunden erprobt wurde, besteht aus einer in die Arterie eingeführten Edelstahlkanüle mit silastischem Überzug, durch die die Blutgase in den analysierenden Teil eines Massenspektrometers geleitet werden, wo ihre Bestimmung erfolgt. Halothan erzeugt eine spektrographische Spitze bei 31, die im Verhältnis zu seinem partiellen Druck steht. Die Spitze bei 31 ergibt sich aus dem Spalten des Halothan-Moleküls in ein Ion, das aus einem Kohlenstoffund einem Fluor-Atom besteht. Die anfängliche Reaktionszeit des Systems beträgt 0,4 Min. Die Zeit vom anfänglichen Übergang auf 63% der Endablesung (1/e Zeit) beträgt 6,3 Min. Die Eichung des Spektrometers blieb bis zu 7 Stunden, dem längsten Versuchszeitraum, stabil.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Brantigan, J. W., Gott, V. L. andMartz, M. N. (1972) A Teflon membrane for measurement of blood and intramyocardial gas tensions by mass spectroscopy.J. Appl. Physiol. 32, 276–282.Google Scholar
  2. Brantigan, J. W., Gott, V. L., Vestal, M. L. Fergusson, C. J. andJohnston, W. H. (1970) A nonthrombogenic diffusion membrane for continuousin vivo measurement of blood gasses by mass spectrometry.J. Appl. Physiol.,28, 375–377.Google Scholar
  3. Papper, E. M. (1970) Uptake of volatile anaesthetic agents in infants and children,in Boulton, T. B., Bryce-Smith, R., Sykes, M. K.,et al. (eds.): Progress in Anaesthesiology, Proceedings of the Fourth World Conference of Anaesthesiology, London 1968, Symposium 13, Amsterdam: Excerpta Medica, 405–406.Google Scholar
  4. Reed, R. R. andOwens, G. (1971) Simultaneous and instantaneous blood flow and tissue gas measurements in experimental stroke therapy.Surgery 70, 254–261.Google Scholar
  5. Roberts, M. andOwens, G. (1972) Mass spectrographic determination of intracerebral Po2 and Pco2.J. Neurosurg. 37, 706–710.CrossRefGoogle Scholar
  6. Roberts, M., Thomas, D. D., Owens, G., Vilinskas, J., andO'Rourke, J. (1972) A new small diameter mass spectrometer cannula for medical use.Med. Res. Eng. 11, 27–28.Google Scholar
  7. Sechzer, P. H., Linde, H. W., Dripps, R. D. andPrice, H. L. (1963) Uptake of halothane by the human body.Anesthesiology,24, 779–783.Google Scholar
  8. Smith, G. V., Ciesel, C. H., Reed, R. R. andOwens, G. (1970) Direct measurement of renal cortical oxygen and carbon dioxide pressures.Surgery,131, 25–28.Google Scholar
  9. Wald, A., Hass, W. K., Siew, F. P. andWood, D. H. (1970) Continuous measurement of blood gassesin vivo by mass spectrography.Med. Biol. Eng.,8, 111–128.Google Scholar
  10. Woldring, S., Owens, G. andWoolford, D. (1966) Blood gases; continuousin vivo recording of partial pressure by mass spectrography.Science,153, 885–887.Google Scholar

Copyright information

© International Federation for Medical & Biological Engineering 1975

Authors and Affiliations

  • Melville Roberts
    • 1
    • 2
  • Ethan T. ColtonIII
    • 1
    • 2
  • Guy Owens
    • 1
    • 2
  • David D. Thomas
    • 1
    • 2
  • George M. Watkins
    • 1
    • 2
  1. 1.University of Connecticut School of MedicineHartfordUSA
  2. 2.Newington Veterans Administration HospitalNewingtonUSA

Personalised recommendations