Medical and biological engineering

, Volume 6, Issue 4, pp 399–402 | Cite as

Left ventricular volume estimation in man using a thermocouple in the ascending aorta

  • B. W. Watson
  • J. S. Fleming
Article

Abstract

A thermocouple is described for the detection of beat by beat temperature changes in the ascending aorta following the injection of cold saline into the left ventricle. Standard manufacturing techniques have been used and the thermocouple can be produced cheaply. The good baseline stability, the frequency response and the linearity of this instrument give it considerable advantages and it can be positioned in the ascending aorta with the minimun of discomfort to the patient. An estimation of the left ventricular end-diastolic volume has been made in 30 patients and no important difficulties were encountered in the use of the thermocouple.

Keywords

Aortic Stenosis Left Ventricular Volume Cold Saline Aorta Ascendens Ventriculaire Gauche 

Sommaire

L'article décrit l'utilisation d'un thermocouple pour la mesure des variations de température, battement aprés battement, dans l'aorteascendante, consécutives à l'injection d'une solution saline froide dans le ventricule gauche. On a utilisé du matériel standard et l'on peut ainsi se procurer les thermocouples à faible prix. La bonne stabilité de la ligne de base, la réponse en fréquence et la linéarité de cet instrument lui donnent des avantages considérables et il peut être placé dans l'aorte ascendante avec le minimum d'inconfort pour le patient. Une estimation du volume ventriculaire gauche à la fin de la diastole a été faite chez 30 sujets et l'on n'a rencontré aucune difficulté dans l'usage du thermocouple.

Zusammenfassung

Eine Thermosonde in der Aorta ascendens zur Messung von Temperaturänderungen zwischen den Herzschlägen nach Injektion kalter Salzlösung in den linken Ventrikel wird beschrieben. Es wurden übliche Herstellungsverfahren verwendet. Die Thermosonde kann billig hergestellt werden. Die gute Stabilität der Basislinie, die Frequenzempfindlichkeit und die Linearität des Geräts stellen beträchtliche Vorteile dar. Die Einführung in die Aorta ascendens ist für den Patienten nur mit geringer Unannehmlichkeit verbunden. Bei 30 Patienten wurde das enddiastolische Volumen im linken Ventrikel bestimmt. Bei der Verwendung der Thermosonde ergaben sich keine wesentlichen Schwierigkeiten.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Brockenbrough, E. C., Braunwald, E. andRoss, J. (1962) Transseptal left heart catheterisation: a review of 450 studies and description of an improved technique.Circulation 25, 15.Google Scholar
  2. Fegler, G. (1954) Measurements of cardiac output in anaesthetised animals by a thermodilution method.Quart. J. exp. Physiol. 39, 153.Google Scholar
  3. Fleming, J. andHamer, J. Left ventricular volume in aortic stenosis measured by an angiocardiographic and a thermodilution method.Br. Heart J. (in press).Google Scholar
  4. Hamer, J. andFleming, J. (1967) The effect of propranolol on left ventricular work in aortic stenosis.Br. Heart J. 29, 871.Google Scholar
  5. Hamilton, W. F., Moore, J. W., Kinsman, J. M. andSpurling, R. G. (1932) Studies on the circulation. Further analysis of the injection method, and of changes in haemodynamics under physiological and pathological conditions.Am. J. Physiol. 99, 534.Google Scholar
  6. Holt, J. P. (1954) Estimation of left ventricular end-diastolic volume and end-systolic volume in the dog. (Abstr.)Am. J. Physiol. 179, 645.Google Scholar
  7. Rolett, E. L., Sherman, K. andGorlin R. (1964) Measurement of left ventricular volume by thermodilution: an appraisal of technical errors.J. appl. Physiol. 19, 1164.Google Scholar
  8. Holt, J. P. (1956) Estimation of the residual volume of the ventricle of the dog's heart by two indicatordilution technics.Circulation Res. 4, 187.Google Scholar

Copyright information

© International Federation for Medical and Biological Engineering 1968

Authors and Affiliations

  • B. W. Watson
    • 1
  • J. S. Fleming
    • 1
  1. 1.Department of Medical Electronics and the Cardiac DepartmentSt. Bartholomew's HospitalLondon