Waveform analysis of surface electrode EMG's used to give independent control signals from adjacent muscles

  • B. H. Brown
Article

Abstract

A technique is described to enable separate EMG control signals to be obtained from surface electrodes placed only 1·5 cm apart over adjacent active muscles. Two EMG signals are recorded with respect to an indifferent electrode and the signals displayed as an x, y plot on an oscilloscope. Different patterns are produced corresponding to different physical actions, as the phase-amplitude relation of the two signals is a function of the position of the contracting muscle. An electronic system is described which enables an output signal to be derived each time the oscilloscope trace is within given pairs of coordinates and by this means signals, which are free from cross-talk, can be obtained from closely spaced surface electrodes.

Keywords

Motor Unit Contracting Muscle Oscilloscope Trace Single Muscle Fibre Indifferent Electrode 

Sommaire

L'article présente une technique de séparation de signaux de commande EMG obtenue à l'aide d'électrodes de surface distantes de 1,5 cm seulement, disposées au-dessus de muscles actifs voisins. Deux signaux d'EMG référés à une même électrode indifferente sont visualisés sur les éntrées x et y d'un oscilloscope. Il apparaît des formes d'onde correspondant à des actions mécaniques différentes, la relation phase-amplitude des deux signaux étant fonction de la position du muscle au travail. Un système électronique dont on trouvera une description, déclenche un signal à chaque fois que la trace de l'oscilloscope se situe à l'intérieur d'un groupe de données bein définies; cela permet d'obtenir des signaux non entachés de diaphonie, à partir d'électrodes très rapprochées.

Zusammenfassung

Eine Methode zur Registrierung getrennter EMG-Kontrollsignale von Oberflächenelektroden, welche im Abstand von nur 1,5 cm über benachbarten aktiven Muskeln angebracht sind, wird beschrieben. Zwei EMG-Signale werden gegen eine indifferente Bezugselektrode aufgenommen. Die Signale werden in kartesischen Koordinaten am Oszillographen sichtbar gemacht. Den verschiedenen Körperbewegungen entsprechen verschiedene Muster, da die Phasen-Amplituden-Beziehung der beiden Signale eine Funktion der Position des kontrahierenden Muskels ist. Ein elektronisches System wird beschrieben, welches die Ableitung eines Ausgangssignals immer dann ermöglicht, wenn die Oszillographenspur innerhalb eines vorgegebenen Koordinatenpaars liegt. Dadurch können unbeeinflußte Signale von nahe benachbarten Oberflächenelektroden erhalten werden.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Bottomley, A. H. andCowell, T. K. (1964) An artificial hand controlled by the nerves.New Scientist. No. 382.Google Scholar
  2. Brown, B. H. (1968) Theoretical and experimental waveform analysis of human nerve action potentials from surface electrodes.Med. biol. Engng 6, 375–386.CrossRefGoogle Scholar
  3. Düring, J. andVan Mittenburg, T. C. M. (1967) An EMG operated control system for a prosthesis.Med. biol. Engng 5, 597–601.CrossRefGoogle Scholar
  4. Ekstedt, J. (1964) Human single muscle fibre action potentials.Acta physiol. scand. 61, Suppl. 226, 96 pp.Google Scholar
  5. Johnson, S. G. (1968) Application of the reverse emitterbase breakdown and saturation of silicon planar transistors.Electron Lett. 4, 109–110.Google Scholar
  6. Nightingale, A. (1957) Ph.D. thesis. Univ. London.Google Scholar
  7. Scott, R. N. (1967) Myo-electric energy spectra.,Med. biol. Engng 5, 303–305.CrossRefGoogle Scholar

Copyright information

© International Federation for Medical and Biological Engineering 1968

Authors and Affiliations

  • B. H. Brown
    • 1
  1. 1.Regional Medical Physics DepartmentSheffieldEngland

Personalised recommendations