Limb-load alarm device for partial-weight-bearing walking exercise

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

A selfcontained electronic device is described for controlling the limb load exerted by a patient in a partial-weight-bearing walking exercise. The device measures the limb load with two removable load transducers attached to the sole of the patient's shoe, compares the load with two preset thresholds and emits audible biofeedback tones. A low-frequency tone is emitted when the load falls in the desired range and a high-frequency tone is activated when the load exceeds the maximum load allowable for the patient. The device has an accuracy of ±10% in all stages of the stance phase. Experimental results that demonstrate the effectiveness of the device in controlling the limb load are also presented.

Sommaire

Le dispositif électronique autonome décrit dans cet article permet de contrôler la charge exercée sur un membre par un patient lors d'exercices de marche avec appui partiel. Le dispositif mesure la charge sur le membre grâce à deux transducteurs de charge enlevables fixés à la semelle de la chaussure du malade; ils comparent cette charge avec deux valeurs-limite présélectionnées et émettent des signaux sonores de biofeedback. Une tonalité grave est émise lorsque la charge entre dans la gamme de valeur désirée, et une tonalité aigüe se déclenche lorsque la charge dépasse la valeur maximale tolérable pour le malade en question. Ce dispositif a une précision de ±10% à tous les stades de la phase de posture. Les résultats expérimentaux démontrant l'efficacité du dispositif pour le contrôle de la charge sur le membre sont également présentés.

Zusammenfassung

Es wird ein unabhängiges Elektronikgerät zur Reglung der von einem Patienten bei einer teilgewichttragenden Gehübung angewandten Gliederbelastung beschrieben. Mit diesem Gerät wird die Gliederbelastung mit zwei abnehmbaren Belastungswandlern gemessen, die an der Schusohle des Patienten befestigt sind, die Belastung mit zwei voreingestellten Schwellwerten vergleichen und hörbare Biorückführgeräusche ausgesenden. Ein Niederfrequenzton wird dann ausgesendet, wenn die Belastung in den gewünschten Bereich fällt. Ein Hochfrequenzton wird dann erzeugt, wenn de Belastung die dem Patienten zumutbare Belastbarkeitsgrenze übersteigt. Das Gerät zeigt während aller Fußstellungsphasen eine Genauigkeit von ±10% an. Die Versuchsresultate, die die Leistung des Geräts bezüglich der Reglung der Gliederbelastung bestätigen, werden auch präsentiert.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Endicott, D., Roemer, R., Brooks, S. andMeisel, H. (1974) Leg load warning system for the orthopaedically handicapped.Med. & Biol. Eng.,12, 318–321.

    Google Scholar 

  • King, P. S., Gerhardt, J. J., Pfeiffer, E. A., Usselman, L. B. andFowlks, E. W. (1972) System for controlling ambulation pressure (SCAP-III) in patients with disabilities of the lower extremity.Amer. J. Phys. Med. 51, 9–15.

    Google Scholar 

  • Miyazaki, S. andIwakura, H. (1978) A foot-force measuring device for clinical assessment of pathological gait.Med. & Biol. Eng. (to be published).

  • Miyazaki, S. (1977) A four-channel PDM/FM telemetry system for continuous measurement of foot-forces during walking.Reports Inst. Med. Dent. Eng.,11, 77–87.

    MathSciNet  Google Scholar 

  • Stolov, W. C., Burgess, E. M. andRomano, R. L. (1971) Progression of weight bearing after immediate prosthesis fitting following below-knee amputation.Arch. Phys. Med. Rehab.,52, 491–502.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Miyazaki, S., Iwakura, H. Limb-load alarm device for partial-weight-bearing walking exercise. Med. Biol. Eng. Comput. 16, 500–506 (1978). https://doi.org/10.1007/BF02457799

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02457799

Keywords

Navigation