Quantitative continuous measurement of pO2 and pCO2 in artery and vein

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

A technique using a quadrupole mass spectrometer has been developed for the continuous measurement of partial pressures of physiological blood gases. The depletion phenomenon is eliminated with a new sampling cannula equiped with a polyethylene membrane for velocities greater than 1 cm per second. The importance of this development resides in the fact that this phenomenon, which is due to a sampling rate around the cannula in excess of the remplacement by the medium of the molecules withdrawn, leads to an under-estimation of the real partial pressure which increases as the blood velocity falls. Thanks to the new cannula, accurate measurements can now be made in both arteries and veins; moreover, continuous observation of physiological phenomena can be carried out with a time constant of under 1 minute resulting from the thinness of the membrane.

Sommaire

La spectrometrie de masse a été adaptée à la mesure continue, quantitative, rapide (constante de temps inférieure à 1 minute) des pressions partielles des gaz dans le sang. Les caractéristiques de la membrane en polyéthylène ainsi que la géométrie de la canule sont telles que la mesure est indépendante de la vitesse du flux sanguin, ce qui permet pour la première fois l’expression en mm Hg des pressions partielles des gaz aussi bien dans le milieu artériel que dans le milieu veineux.

Zusammenfassung

In einem früher veröffentlichten Artikel (Seylaz u.a., 1974) beschrieben wir technische Verbesserungen, die im Vergleich mit anderen bekannten Verfahren zu einer Verringerung der sogenannten Verarmungserscheinung führten. Diese Verbesserung beruht auf der Entwicklung einer asymmetrischen Kanüle und einer silastischen Membran, die so in der Arterie angeordnet werden können, daß die sensitiven Saugzonen dem Lumen zugewandt sind, wo der Blutdurchfluß am schnellsten ist. Mit dieser asymetrischen Kanüle wurde die Verarmung in der Arterie auf ein geringfügiges maß beschränkt und in der Vene, wo das Blut viel langsamer fließt, auf 10–15% herabgesetzt. In diesem Artikel berichten wir über die weiters Entwicklung der Probekanüle, welche, durch Verinfachung der Meßtechnik, die Verarmungserscheinung für schneller als 1 cm pro Sekunde fließendes Blut ganz ausschaltet. Dies führt zu einer höheren Zuverlässigkeit der Messungen, so daß quantitative Messungen in einem beliebigen Blutmedium, einschließlich der Vene, durchgeführt werden können und doch eine gute Zeitkonstante beibehalten wird.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Brantigan, J. W., Gott, V. L., Vestal, M. L., Fergusson, G. J. andJohnston, W. H. (1970). A non thrombogenic diffusion membrane for continuousin vivo measurement of blood gases by mass spectrometry.J. Appl. Physiol. 28, 375.

    Google Scholar 

  • Brantigan, J. W., Gott, V. L. andMartz, M. N. (1972). A Teflon membrane for measurement of blood and intramyocardial gas tensions by mass spectrometry.32, 276–282.

    Google Scholar 

  • Brantigan, J. W., Dunn, K. L. andAlbo, D. (1976). A clinical catheter for continuous blood gas measurement by mass spectrometry.40, 443–446.

    Google Scholar 

  • Delpy, D. andParker, D. (1976). A flexible catheter for continuous measurement of blood gas tensionsin vivo by mass spectrometry. 11th international conference on medical and biological engineering. Ottawa.

  • Grode, G. A., Anderson, S. J. andFalb, R. D. (1969). Non-thrombogenic materials via a simple coating process.Trans. Amer. Soc. Artificial Internal Organs. 15, 1–6.

    Google Scholar 

  • Hass, W. K., Siew, F. P. andYee, D. J. (1968). Progress in adaptation of mass spectrometer to study of human cerebral blood flow.Circulation 38, 96.

    Google Scholar 

  • Key, A. (1975) A flexible catheter for continuous measurement of blood gas tensons.Med & Biol. Eng. 13, 583.

    Google Scholar 

  • Owens, G., Belmusto, L. andWoldring, S. (1969). Experimental intracerebral pO2 and pCO2 monitoring by mass spectrography.J. Neurosurg. 30, 110–115.

    Article  Google Scholar 

  • Seylaz, J., Pinard, E., Correze, J. L., Luft, A. andMamo, H. (1974). Quantitative continuous measurement of blood gas tensions by mass spectrometry.J. Appl. Physiol. 6, 937–941.

    Google Scholar 

  • Wald, A., Hass, W. K., Siew, F. P. andWood, D. W. (1970). Continuous measurement of blood gases “in vivo” by mass spectrography.Med. & Biol. Eng. 8, 111–128.

    Google Scholar 

  • Wald, A., Hass, W. K. andRansohoff, J. (1971). Experience with a mass spectrometer system for blood gas analysis in humans.J. Assn. Adven. Med. Instrum. 5, 325–342.

    Google Scholar 

  • Woldring, S. (1970). Tutorial: biomedical application of mass spectrometry for monitoring partial pressures.J. Assn. Advan. Med. Instrum. 4, 43–56.

    Google Scholar 

  • Woldring, S., Owens, G. andWoolford, D. C. (1966). Blood gases: continuousin vivo recording of partial pressure by mass spectrometry.Science 153, 885–887.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pinard, E., Seylaz, J. & Mamo, H. Quantitative continuous measurement of pO2 and pCO2 in artery and vein. Med. Biol. Eng. Comput. 16, 59–64 (1978). https://doi.org/10.1007/BF02442933

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02442933

Keywords

Navigation