Transventricular defibrillation thresholds using quarter and half sinusoidal pulses

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

Transventricular peak current defibrillation thresholds using quarter (5 ms) and half (10 ms) sinusoidal pulses were determined in dogs under conditions of normothermia. For 5 ms, the overall average was 59·6 mAg−1 of heart (s.d. 21·00) in a total of ten animals and 88 successful defibrillations. For 10 ms, the average over seven dogs and 51 determinations was 52·5 mAg−1 of heart (s.d. 15·4). This difference was not statistically significant, whereas the energy delivered in the latter was about twice as much as in the former, suggesting the quarter sine wave pulse was better for defibrillation. A tighter definition of threshold is also introduced, i.e. the midrange between the maximum value which did not defibrillate and the minimum that did defibrillate. Such a definition is more realistic because it takes into account the successful and also the unsuccessfulvalues within a band of overlapping data points. The thresholds so obtained (52·3 mAg−1, s.d. 15·7, nine dogs, 5 ms; 42·9 mAg−1, s.d. 6·56, fourdogs, 10 ms) were, as expected, somewhat lower than the above mentioned ones showing, at the same time, lower coefficients of variation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Armayor, M. R. Savino, G. V., Valentinuzzi, M. E., Clavin, O. E., Monzon, J. E. andArredondo, M. T. (1979) Ventricular defibrillation thresholds with capacitor discharge.Med. & Biol. Eng. & Comput.,17, 435–442.

    Google Scholar 

  • Bourland, J. D., Tacker, W. A. andGeddes, L. A. (1978a) Strength-duration curves for trapezoidal waveforms of various tilts for tranchest defibrillation in animals.Medical Instrumentation,12, 38–41.

    Google Scholar 

  • Bourland, J. D., Tacker, W. A., Geddes, L. A. andChafee, V. (1978b) Comparative efficacy of damped sine wave and square wave current for tranchest ventricular defibrillation in animals.,12, 42–45.

    Google Scholar 

  • Carlino, A. F., Valentinuzzi, M. E., Torres, J. H. andArredondo, M. T. (1974) Estimulador biológico con unidad aisladora de audiofrecuencia.La Semana Médica (Buenos Aires, Argentina),145, 359–374, 380. AlsoRevista Instituto Ingenieria Eléctrica Universidad Nacional de Tucumán (Tucumán, Argentina),11, 5–22.

    Google Scholar 

  • Clavin, O. E. andValentinuzzi, M. E. (1977) Impedancimetro biológico de alta linealidad.Acta Physiologica Latinoamericana (Buenos Aires, Argentina),27, 215–230.

    Google Scholar 

  • Geddes, L. A., Tacker, W. A. andMcFarlane, J. R. (1973) Ventricular defibrillation with single and twin pulses of half-sinusoidal current.J. Appl. Physiol.,34, 8–11.

    Google Scholar 

  • Koss, M. R., Valentinuzzi, M. E. andEscalante Figueroa, D. (1977) Respirador por presión positiva controlado electrónicamente.Mundo Electrónico (Spain),58, 57–61.

    Google Scholar 

  • McFarlane, J. R., Geddes, L. A., Milnor, W., Tacker, W. A., Bourland, J. andCoulter, T. W. (1971) Ventricular defibrillation with single and multiple half sinusoidal pulses of current.Cardiovasc. Res.,5, 286–292.

    Article  Google Scholar 

  • Monzón, J. E., Guillen, S. G. andValentinuzzi, M. E. (1980) Desfibrilador por pulsos senoidales.Mundo Electrónico (Spain),93, 37–40.

    Google Scholar 

  • Moore, T. W., di Meo, F. N. andDubin, S. E. (1977a) Ventricular defibrillation using optimally timed staggered shocks.Digest First Mediterranean Conf. Med. & Biol. Eng.,II, September, 10–25/10–28.

    Google Scholar 

  • Moore, T. W., di Meo, F. N. andDubin, S. E. (1977b) The half-cycle sinusoid as an alternative defibrillating waveform in low-energy applications.Annals Biomed. Eng.,5, 157–163.

    Article  Google Scholar 

  • Savino, G. V., Ruiz, E. andValentinuzzi, M. E. (1981) Two methods for estimating transventricular impedance during fibrillation.Vth International Conference on Electrical Bio-Impedance, August 24–26, Tokyo, Japan.

  • Tacker, W. A. andGeddes, L. A. (1980)Electrical Defibrillation. CRC Press, Boca Raton, Florida, USA.

    Google Scholar 

  • Tacker, W. A., Geddes, L. A. andBourland, J. D. (1971) Defibrillation of the dog ventricles using single and multiple half-sinusoidal current pulses.Cardiovasc. Res. Center. Bull.,10, 57–67.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Monzón, J.E., Valentinuzzi, M.E. Transventricular defibrillation thresholds using quarter and half sinusoidal pulses. Med. Biol. Eng. Comput. 20, 756–760 (1982). https://doi.org/10.1007/BF02442532

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02442532

Key words

Navigation