Sterile production and mechanical reliability of autoclaves

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

The failure rate of a steam autoclave used in a hospital sterile-fluid production unit has been analysed by fitting observed breakdowns to the Weibull distribution. The results indicate a rapid initial breakdown rate which clearly represents an unsatisfactory situation for the pharmacy department trying to provide a consistent, reliable service. It is concluded that greater attention should be paid to auctoclave reliability if cost-effective sterile-fluid production is to be achieved

Sommaire

Le taux d'échecs d'un autoclave à vapeur, employé dans un hôpital pour la production de liquide de stérilisation, a été analysé en arrangeant des pannes à la distribution Weibull. Les résultats indiquent rapidement un taux de pannes, ce qui constitue clairement une situation insatisfaisante pour les services pharmaceutiques, essayant de fournir un service régulier et sûr. Conclusion: il faut donc apporter une plus grande attention à la fiabilité de l'autoclave, si l'on veut réaliser une production de liquide de stérilisation à un prix raisonnable.

Zusammenfassung

Die Ausfallrate eines in einer sterilen Flüssigkeitsproduktionseinheit eines Krankenhauses benutzten Dampfautoklaven wurde durch Einfügen beobachteter Versagen in die Weibull-Distribution analysiert. Die Resultate weisen auf einschnelles erstmaliges Versagen hin, was deutlich eine unzufriedenstellende Situation für die pharmazeutische Abteilung darstellt, die versucht, einen fortlaufend zuverlässigen Service zu bieten. Man kam zu der Übereinstimmung, daß der Zuverlässigkeit der Autoklaven mehr Beachtung geschenkt werden muß, wenn eine kosteneffektive, sterile Flüssigkeitsproduktion erreicht werden soll.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Bompas-Smith, J. H. (1973)Mechanical Survival: the use of reliability data. McGraw-Hill, London.

    Google Scholar 

  • Hoskins, H. T. (1976) Failure rate of hospital spray-cooled autoclaves. 1976 Merck, Sharp and Dohme Award for the advancement of hospital pharmacy.

  • Lees, F. P. (1976) The reliability of instrumentation.Chemistry and Industry, 6th March, 195.

  • Peto, R., Pike, M. C., Armitage, P., Breslow, N. E., Cox, D. R., Howard, S. V., Mantel, N., McPherson, K., Peto, J. andSmith, P. G. (1977) Design and analysis of randomized clinical trials requiring prolonged observation of each patient.Br. J. Cancer,35, 1–39.

    Google Scholar 

  • Sherwin, D. J. (1976) Personal communication.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hoskins, H.T., Diffey, B.L. Sterile production and mechanical reliability of autoclaves. Med. Biol. Eng. Comput. 16, 330–333 (1978). https://doi.org/10.1007/BF02442436

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02442436

Keywords

Navigation