Construction and performance of a new catheter-tip oxygen electrode

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

A membrane-covered catheter-tip oxygen-electrode system is described, which gives a linear response in the Po2 range of 0–350 mm Hg. The system is highly stable, free from drift and mechanically safe for application in man. This is accomplished by using a screw cap for fastening the membrane holder, thus preventing the loss of parts and making the electrolyte chamber really fluidtight. Insulation of the platinum wire with glass precludes the possibility of fluid-bridge contact with the silver anode beyond the measuring site at the tip.

Sommaire

Est décrit un système d'électrode à oxygène munie d'un embout cathéter et recouverte d'une, membrane, donnant une réponse linéaire dans la gammeP o2 de O à 350 mm Hg. Ce système est extrêmement stable, exempt de dérive et du point de vue mécanique il ne présente aucun danger d'application sur l'homme. Une telle sûreté d'application est obteneue en utilisant un capuchon à vis pour fixer le support de membrane, permettant ainsi d'éviter la perte de pièces et assurant l'étanchéité complète de la chambre électrolytique. Grâce au verre isolante le fil de platine, toute possibilité de contact entre le fluide et l'anode d'argent au-delà du niveau de mesure de l'embout est exclue.

Zusammenfassung

Beschrieben wird ein membranumhulltes Sauerstoffelektrodensystem mit Katheterspitze, das imP o2-Bereich von 0–350 mm Hg eine Linearreaktion aufweist. Das System ist mechanisch sehr stabil, frei von Nullpunktverschiebung und kann gefahrlos im Menschen angewendet werden. Dies wird durch Verwendung einer Schraubenkappe zur Befestigung des Membranhalters ermöglicht, was das Verlorengehen von Bestandteilen ausschaltet und die Elektrolytenzelle wirklich flüssigkeitsdicht macht. Glasisolierung des Platindrahtes schließt die Möglichkeit eines Flüssigkeitsbrückenkontaktes mit der Silberanode außerhalb der Meßstelle an der Spitze aus.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Berkenbosch, A. (1967) Time course of response of the membrane-covered oxygen electrode.Acta Physiol. Pharmacol. Neerl.,14, 300–316.

    Google Scholar 

  • Clark, L. C., Jr. (1956) Monitor and control of blood and tissue oxygen tension.Trans. Am. Soc. Artif. Int. Organs,2, 41–45.

    Google Scholar 

  • Eberhard, P., Mindt, W., Jann, F. andHammacher, K. (1975) ContinuousP o2 monitoring in the neonate by skin electrodes.Med. & Biol. Eng.,13, 436–442.

    Google Scholar 

  • Heller, M. L., Imredy, D. S., Shafer, P. R. andWatson T. R. (1972)A P o2 Catheter microelectrode forin vivo intravascular use.Anesthesia & Analgesia,51, 453–461.

    Google Scholar 

  • Huch, A., Huch, R. andRooth, G. (1973a) Monitoring the intravascularP o2 in newborn infants.J. Perinat. Med.,1, 53–59.

    Google Scholar 

  • Huch, R., Lübbers, D. W. andHuch, A. (1973b) Transcutaneous measurement of bloodP o2.J. Perinat. Med.,1, 183–186.

    Article  Google Scholar 

  • Key, A., Parker, D. andDavies, R. (1970) Use of epoxy resin in oxygen electrodes.Phys. Med. Biol.,15, 569–572.

    Article  Google Scholar 

  • Kornel, J., Hemptinne, J. de, Swietochowski, A. andDemeester, M. (1975) Continuousin vivo measurement of arterialP o2 in humans.J. Appl. Physiol.,38, 730–735.

    Google Scholar 

  • Oeseburg, B., Landsman, M. L. J., Mook, G. A. andZijlstra, W. G. (1972) Direct recording of oxyhaemoglobin disscciation curvein vivo.Nature,237, 149–150.

    Article  Google Scholar 

  • Siu, W., andCobbold, R. S. G. (1976) Characteristics of a multicathode polarographic oxygen electrode.Med. & Biol. Eng.,14, 109–121.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Jansen, T.C., Lafeber, H.N., Visser, H.K.A. et al. Construction and performance of a new catheter-tip oxygen electrode. Med. Biol. Eng. Comput. 16, 274–277 (1978). https://doi.org/10.1007/BF02442427

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02442427

Keywords

Navigation