Skip to main content
Log in

Comparison of scintigraphic count-volume and geometric methods for measuring left ventricular ejection fraction

  • Published:
Annals of Biomedical Engineering Aims and scope Submit manuscript

Abstract

A comparison of geometric and count-volume scintigraphic methods for measuring left ventricular ejection fraction (EF) is presented. The particular count-volume method used is a modification of an established first-pass technique. It is shown that a modification of the technique is necessary to prevent a positive bias in the EF estimate due to Poisson noise. Using the same simulated data, EF estimates by the modified technique were always closer to the actual EF volume than were the estimates by the original method. Maximum Likelihood estimates were also more accurate than were the other two estimates. In 56 patients with either acute myocardial infarcts or unstable angina pectoris, the scintigraphic single-plane gated geometric method tended to overestimate EF in each of the patient groups when compared with the geometry-independent count-volume estimates. The results also point out that the error in the geometric estimates tended to be greater for larger ejection fractions and appears to be due to the systematic underestimation of end-systolic volumes.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Chapman, C. B., Baker, O., Reynolds, J., and Bonte, F. J., Use of biplane cinefluorography for measurement of ventricular volume,Circulation 1958,18, 1105.

    PubMed  CAS  Google Scholar 

  • Chapman, C. B., Baker, V., Mitchell, J. H., and Collier, R. G., Experiences with a cinefluoro-graphic method for measuring ventricular volume.American Journal of Cardiology 1966,18, 25–30.

    PubMed  CAS  Google Scholar 

  • Davila, J. C., and Sanmarco, M. E., An analysis of the fit of mathematical models applicable to the measurement of left ventricular volume.American Journal of Cardiology 1966,18, 31–42.

    PubMed  CAS  Google Scholar 

  • Dodge, H. R., Sandler, H., Ballew, D. W., and Lord, J. D., Jr., The use of biplane angiocardiography for measurement of left ventricular volume in man,American Heart Journal 1960,60, 762.

    PubMed  CAS  Google Scholar 

  • Keshavan, H. R., Nardizzi, L. R., Srinath, M. D., and Doss, J., Noninvasive radioisotope cardiographic images with digital prefiltering and boundary detection for measuring left ventricular parameters,Mathematical Biosciences, 1977,37, 223–243.

    Article  Google Scholar 

  • Melsa, J. L., and Sage, A. P.,An introduction to probability and stochastic processes, Englewood Cliffs, N.J.: Prentice-Hall, 1973.

    Google Scholar 

  • Mitchell, J. H., Wildenthal, K., and Mullins, C. B., Geometrical studies of the left ventricle utilizing biplane cinefluorography,Federation Proceedings 1969,28, 1334–1343.

    PubMed  CAS  Google Scholar 

  • Mullins, C. B., Mason, D. T., Ashburn, W. L., et al., Determination of ventricular volume by radioisotope angiography,American Journal of Cardiology 1969,24, 72–78.

    Article  PubMed  CAS  Google Scholar 

  • Rackley, C. E., Quantitative evaluation of left ventricular function by radiographic techniques,Circulation 1976,54, 862–879.

    PubMed  CAS  Google Scholar 

  • Rigo, P., Murray, M., Strauss, H. W., Taylor, D., Kelley, D., Weisfeldt, M., and Pitt, B., Left ventricular function in acute myocardial infarction evaluated by gated scintiphotography,Circulation 1974,50, 678.

    PubMed  CAS  Google Scholar 

  • Sandler, H., Dimensional analysis of the heart—A review,American Journal of Medical Sciences 1970,260, 56–70.

    CAS  Google Scholar 

  • Sandler, H., and Dodge, H. F., The use of single plane angiocardiograms for the calculation of left ventricular volume in man,American Heart Journal 1968,75, 325.

    Article  PubMed  CAS  Google Scholar 

  • Schelbert, H. R., Verba, J. W., Johnson, A. D., Brock, G. W., Alazraki, N. P., Rose, F. J., and Ashburn, W. L., Nontraumatic determination of left ventricular ejection fraction by radionuclide angiography,Circulation 1975,51, 902–909.

    PubMed  CAS  Google Scholar 

  • Snedecor, G. W., and Cochran, W. G.,Statistical methods, Ames, Iowa: Iowa State Univ. Press, 1967.

    Google Scholar 

  • Stokely, E. M., Parkey, R. W., Bonte, F. J., Graham, K. D., Stone, J., and Willerson, J. T., Gated blood pool imaging following99mTc stannous prophosphate imaging,Radiology 1976,120, 433.

    PubMed  CAS  Google Scholar 

  • Twieg, D. B., Stokely, E. M., Parkey, R. W., Doss, J. K., Pulido, J., Poliner, L., Bonte, F. J., and Willerson, J. T., Noninvasive myocardial imaging to define myocardial performance in patients with acute ischemic heart disease (Abstract),Clinical Research 1977,25.

Download references

Author information

Authors and Affiliations

Authors

Additional information

This work was supported in part by NSF Grant ENG 73–03789 A02, by NIH Ischemic SCOR Grant HL-17669, and by NIH Grant HL-17777.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Twieg, D., Nardizzi, L.R., Parkey, R. et al. Comparison of scintigraphic count-volume and geometric methods for measuring left ventricular ejection fraction. Ann Biomed Eng 6, 310–325 (1978). https://doi.org/10.1007/BF02409347

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02409347

Keywords