, Volume 77, Issue 2, pp 623–628 | Cite as

Mass spectrometry, with an inductively coupled plasma, of inorganic trace impurities in ultrapure water

  • G. I. Kir'yanov
  • V. V. Titov
Articles

Conclusions

The final result of investigations of the mass spectrometer with an inductively coupled plasma for analysis of inorganic trace impurities in ultrapure water was the development of instrumental-methodological techniques for increasing the efficiency and reliability of the recording process. This made it possible to choose internal calibration standards and criteria for changing the calibration coefficients. The research instrumentation complex, developed at the All-Union Scientific-Research Institute of Technical and Physical Apparatus, is suitable for quantitative analysis of the elemental content of trace impurities in water with the indicated analytic characteristics for the relative method of measurement according to internal calibration standards. A further improvement of the research apparatus is lower transmission losses which result from the effect of the edge fields according to the constant and variable components, with the aid of a nonuniform standing wave of voltage applied to the mass analyzer as well as the detection limit obtained with the aid of an ion-trap-type detector. The preliminary results show that these techniques make it possible to improve transmission by up to a factor of 30 and to work with a resolution of 100M, as well as to decrease the detection limits by a factor of 10.

In conclusion, we thank S. S. Grazhulen and Yu. I. Popandopulo (Institute of Applied and Theoretical Mechanics of the Russian Academy of Sciences) for collaboration and assistance in performing the control measurements of the impurity content in the samples on the VG Elemental PQ2 mass spectrometer with an inductively coupled plasma, S. G. Galaktionov and A. L. Makarov (All Union Scientific-Research Institute of Chemical Technology) for providing the samples, and A. A. Sysoev (Moscow Engineering Institute) for a helpful discussion of the results.

Keywords

Ultrapure Water Standing Wave Couple Plasma Impurity Content Transmission Loss 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Literature Cited

  1. 1.
    R. Hutton, M. Bridenne, E. Coffre, et al., "Investigations into the direct analysis of semiconductor grade gases by inductively coupled plasma mass spectrometry," J. Anal. Atomic Spectrom.,5, 463–470 (1990).CrossRefGoogle Scholar
  2. 2.
    G. I. Ramendik, D. A. Tjurin, and Yu. I. Babikov, "Is a universal model for ion formation during mass spectrometric elemental analysis possible?" Anal. Chem.,62, 2501–2503 (1990).CrossRefGoogle Scholar
  3. 3.
    V. V. Titov, "Energy of ions in the distorted fields of a quadrupole mass analyzer," At. Énerg.,76, No. 3, 219–226 (1994).Google Scholar
  4. 4.
    V. V. Titov, "Theoretical aspects of the operation of a quadrupole mass spectrometer with a distorted analyzing field," At. Énerg.,61, No. 2, 103–110 (1986).Google Scholar
  5. 5.
    V. V. Titov, "Analytic and numerical methods for modeling ion-optic systems of a quadrupole mass spectrometer," At. Énerg.,75, No. 2, No. 8, 109–121 (1993).Google Scholar
  6. 6.
    V. V. Titov, "Experimental aspects of ion separation in the distorted fields of a quadrupole mass analyzer," Voprosy Atomnoi Nauki i Tekhniki. Ser. Tekhnicheskaya Fizika i Avtomatizatsiya, No. 1/2(49/50), 35–49 (1994).Google Scholar
  7. 7.
    V. V. Titov, "Isotopic quadrupole mass spectrometry of hydrogen and helium," J. Radioanal. Nucl. Chem. Art.,174, No. 2, 205–222 (1993).CrossRefGoogle Scholar
  8. 8.
    G. I. Kir'yanov and V. V. Titov, "Determination of inorganic impurities in high pure water by ICP-MS," ICP Inf. Newsletter,18, No. 7, 412 (1992).Google Scholar

Copyright information

© Plenum Publishing Corporation 1995

Authors and Affiliations

  • G. I. Kir'yanov
  • V. V. Titov

There are no affiliations available

Personalised recommendations