Summary

A method enabling a fully automated computer analysis of the visual field is described. Starting from the most probable assumptions about the sensitivity distribution within the visual field, this distribution is approximated in 4 steps.

Decisions are based on probability theory. Every analytical step builds up on the conclusions reached in the preceding one. The theoretical limits of the method in regard to the degree of approximation of the true sensitivity function are discussed in detail. The spatial resolution is determined by the density of the questions per area of visual field. For a grid constant of 3°, the spatial resolution of sensitivity defects varies between 4 and 7°.

Otherwise, the quality of the approximation varies within large limits and depends on prior knowledge of the expected sensitivity distribution and on the noise in the system. This includes the patient's threshold fluctuation and the reliability of the answers.

The whole analytical program was tested on a large number of artificial visual fields contained in computer storage which were then tested by the main program. It is shown that even in the presence of large sensitivity fluctuations and a considerable fraction of erroneous answers by the patient the method is still able to extract the data which are essential from the clinical point of view.

Zusammenfassung

Es wird eine Methode der vollautomatisierten Perimetrie beschrieben. Die Empfindlichkeitsverteilung im Gesichtsfeld wird, ausgehend von der wahrscheinlichsten Annahme, in 4 Schritten angenähert. Den Entscheiduugen liegen wahrscheinlichkeitstheoretische Überlegungen zugrunde, wobei jeder analytische Schritt auf den Schlüssen des vorausgehenden aufbaut. Die theoretischen Grenzen der Methode hinsichtlich der Approximationsgüte werden in allen Einzelheiten geschildert. Die räumliche Auflösung hängt von der Dichte der Testpunkte im Gesichtsfeld ab. Bei einer Gitterkonstanten von 3° schwankt das räumliche Auflösungsvermögen für Scotome zwischen 4 und 7°. Die Approximationsgüte schwankt in weiten Grenzen. Sie hängt von der Vorkenntnis der Empfindlichkeitsverteilung im Gesichtsfeld und dem Rauschen im neuro-visuellen System ab. Schwankungen der Empfindlichkeitsschwelle und Konzentrationsstörungen des Patienten, die zu falschen Antworten führen, setzen die Approximationsgüte herab.

Das ganze analytische Programm wurde an einer großen Zahl künstlicher Gesichtsfelder geprüft, die im Computer gespeichert wurden. Dabei zeigte sich, daß die Methode selbst bei einer hohen Patientenstreuung und einem hohen Prozentsatz falscher Antworten infolge von Konzentrationsstörungen noch in der Lage ist, aus einem komplexen Gesichtsfeld die Informationen zu extrahieren, die vom klinischen Standpunkt aus von Bedeutung sind.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literature

  • Aspinall, P. A.: Variables affecting the retinal threshold gradient in static perimetry. Master of Science Thesis. Department of Psychology. University of Edinburgh, June 1967.

  • Békésy, G. von: A new audiometer. Acta oto-laryng. (Stockh.)35, 411–422 (1947).

    Google Scholar 

  • Cornsweet, T. N.: The staircase method in psychophysics. Amer. J. Psychol.75, 485–491 (1962).

    CAS  PubMed  Google Scholar 

  • Fankhauser, F., Schmidt, T.: Die optimalen Bedingungen für die Untersuchung der r⇓umlichen Summation mit stehender Reizmarke nach der Methode der quantitativen Lichtsinnperimetrie. Ophthalmologica (Basel)139, 409–423 (1960).

    CAS  Google Scholar 

  • Goldmann, H.: Demonstration unseres neuen Projektionskugelperimeters samt theoretischen und klinischen Bemerkungen über Perimetrie. Ophthalmologica (Basel)111, 187–192 (1946).

    Google Scholar 

  • Hüsser, R.: Orthogonale Polynome mehrerer Veränderlichen und ihre Anwendung in der ein-und zweidimensionalen Ausgleichsrechnung. Inaugural-Dissertation der philosophisch-naturwissenschaftlichen Fakultät der Universität Bern 1957.

  • Hüsser, R., Wegmüller, W.: Einsatz elektronischer Rechenautomaten für Ausgleichung mit orthogonalen Polynomen. Vereinigung schweiz. Versicherungsmathematiker60, 151–169 (1960).

    Google Scholar 

  • Koch, P., Roulier, A., Fankhauser, F.: Perimetry—the information theoretical basis for its automation. Vision Res. in print.

  • Lynn, J. R.: Personal communication 1970.

  • Sunga, R. N., Enoch, J. M.: Further perimetric analysis of patients with lesions of the visual pathway. Amer. J. Ophthal.70, 403–422 (1970).

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

This work was supported in part by a Grant to the University of Bern, Number EY 00232 from the NINDB, National Institutes of Health, Bethesda, Maryland, USA, and in part by the Schweizerischer Nationalfonds zur Förderung der wissenschaftlichen Forschung.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Fankhauser, F., Koch, P. & Roulier, A. On automation of perimetry. Albrecht v Graefes Arch. klin. exp. Ophthal. 184, 126–150 (1972). https://doi.org/10.1007/BF02390260

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02390260

Keywords

Navigation