Abstract
The operation and maintenance of a complex analytical instrument such as an inductively coupled plasma mass spectrometer in a radioactive contaminated environment presents unique problems and challenges that have to be considered in the purchasing and installation process. Considerations such as vendor experience, typical radiation levels, sample matrices encountered during sample analysis, instrument accessability for maintenance, and upkeep must be incorporated into the decision process. The Radioactive Materials Analytical Laboratory (RMAL) at Oak Ridge National Laboratory (ORNL) recently purchased and installed an inductively coupled plasma mass spectrometer for the analysis of Department of Energy (DOE) radioactive waste streams. This presentation will outline the purchasing decision, installation of the instrument, and how the modifications needed to operate in a radioactive contaminated laboratory do not significantly impact the daily operation and maintenance requirements of the instrument. Also, a contamination survey of the system will be presented which demonstrates the contamination levels in the instrument from the sample introduction system to the detector.
Similar content being viewed by others
References
J. M. Giaquinto, J. M. Keller, T. P. Mills, Miscellaneous Data for the 1996–1997 Sampling and Analysis Campaigns of the MVST, BVEST, and OHF Tank Complexes, ORNL/TM-13455, Oak Ridge National Laboratory, 1997.
J. M. Keller, J. M. Giaquinto, A. M. Meeks, Characterization of the MVST Waste Tanks Located at ORNL, ORNL/TM-13357, Oak Ridge National Laboratory, 1996.
J. M. Keller, J. M. Giaquinto, A. M. Meeks, Characterization of the BVEST Waste Tanks Located at ORNL, ORNL/TM-13358, Oak Ridge National Laboratory, 1997.
J. M. Keller, J. M. Giaquinto, A. M. Meeks, Characterization of the Old Hydrofracture Facility (OHF) Waste Tanks Located at ORNL, ORNL/TM-13394, Oak Ridge National Laboratory, 1997.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Giaquinto, J.M., Keller, J.M. & Meeks, A.M. Inductively coupled plasma mass spectrometer installation modifications in a radioactive contaminated laboratory for the analysis of DOE radioactive waste streams. J Radioanal Nucl Chem 234, 137–141 (1998). https://doi.org/10.1007/BF02389761
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02389761