Experimental and numerical studies on convective heat transfer in a neonatal incubator

  • Y. -H. Kim
  • C. -H. Kwon
  • S. -C. Yoo
Article

Abstract

Thermo-neutrality is one of the major environmental factors affecting a premature or low-birth-weight neonate inside an incubator. Severe temperature differences inside an incubator lead to neonate heat loss, hypothermia and apnoea, which are closely related to air flow and air velocity. In the study, flow visualisations, hot-wire velocity measurements and computational fluid dynamics simulate the airflow inside a neonatal incubator. An anatomically correct neonate model is designed using a three-dimensional laser scanner system and a rapid prototyping machine. Flow visualisations demonstrate that large-scale rotating airflow is produced inside the chamber, and a number of small, stationary eddies are found in regions between the air inlet and the neonate. Hot-wire measurements show that air velocities along the long inlets are not uniform. Computational fluid dynamics show relatively uniform temperatures of about 34°C on the neonate's anterior aspect and the highest temperature of 36.1°C at the right armpit and the crotch. Flow fields from airflow visualisations, hot-wire measurements and computational fluid dynamics are very similar, both qualitatively and quantitatively. The small eddies produced between the neonate and the mattress could interfere with convective and evaporative heat transfers from the neonate. Therefore it is important to eliminate eddies around the neonate in future designs of neonatal incubators.

Keywords

Neonatal incubator Flow visualisations Hot-wire anemometry Rotating vortices Turbulence 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Bell, E. F. (1983): ‘Infant incubators and radiant warmers’,Early Hum. Dev.,8, pp. 351–375CrossRefGoogle Scholar
  2. Bell, E. F., andRios, G. R. (1983): ‘A double-walled incubator alters the partition of body heat loss of premature infants’,Pediatr. Res.,17, pp. 135–140Google Scholar
  3. Faranoff, A. A., Wald, M., Gruber, H. S., andKraus, M. H. (1972): ‘Insensible water loss in low birth weight infants’,Pediatrics,50, pp. 236–245Google Scholar
  4. Hasegawa, T., Horio, H., Okino, H., Taylor, T. W., andYamaguchi, T. (1993): ‘Three-dimensional visualization of air flow in infant incubators using computational fluid mechanics’,Biomed. Instrum. Technol., pp. 311–317Google Scholar
  5. Hey, E. N. (1966): ‘Temperature control in incubators’,Lancet,2, p. 202Google Scholar
  6. Hey E. N. (1969): ‘The relation between environmental temperature and oxygen consumption in the newborn baby’,J. Physiol. 200, pp. 589–603Google Scholar
  7. Hey, E. N., andKatz, G. (1970): ‘The optimum thermal environment for naked babies’,Arch. Dis. Child.,45, pp. 328–334Google Scholar
  8. Hey, E. N., andMount, L. E. (1972): ‘Heat losses from babies in incubators’,Arch. Dis. Child.,42, pp. 75–84Google Scholar
  9. Johnson, A. T. (1991): ‘Biomechanics and exercise physiology’ (Wiley & Sons), pp. 361–368Google Scholar
  10. Levene, M. I., Tudehope, D. I., andThearle, M. J. (2000): ‘Essentials of neonatal medicine, 3rd edn’ (Blackwell Science, 2000), pp. 89–90Google Scholar
  11. Libert, J. P., Bach, V., andFarges, G. (1997): ‘Neural temperature range in incubators: performance of equipment in current use and new developments’,Crit. Rev. Biomed. Eng.,25, pp. 287–370Google Scholar
  12. Okken, A., Blijham, C., Franz, W., andBohn, E. (1982): ‘Effects of forced convection of heated air on insensible water loss and heat loss in preterm infants in incubators’,J. Pediatr.,101, pp. 108–112Google Scholar
  13. Perlstein, P. H., Edwards, N. K., Atherton H. D., andSutherland, J. M. (1976): ‘Computer-assisted newborn intensive care’,Pediatrics,57, pp. 494–502Google Scholar
  14. Perlstein, P. H., Edwards, N. K., andSutherland, J. M. (1970): ‘Apnea in premature infants and incubator-air-temperature changes’,N. Engl. J. Med.,282, pp. 461–466Google Scholar
  15. Sarman, I., Bolin, A., Holmer, I., andTunell, R. (1992): ‘Assessment of thermal conditions in neonatal care: use of a manikin of premature baby size’,Am. J. Perinatol.,9, pp. 239–246Google Scholar
  16. Stothers, J. K. (1980): ‘The effect of forced convection on neonatal heat loss’,J. Physiol.,305, p. 77Google Scholar
  17. Stothers, J. K., andWarner, R. M. (1984): ‘Thermal balance and sleep state in the newborn’,Early Hum. Dev.,9, pp. 313–322CrossRefGoogle Scholar
  18. Telliez, F., Bach, V., Delanand, S., Bouferrache, B., Krim, G., andLibert, J. P. (1997): ‘Skin derivative control of thermal environment in a closed incubator’,Med. Biol. Eng. Comput.,35, pp. 521–527Google Scholar
  19. Thompson, M. H., Stothers, J. K., andMcLellan, N. J. (1984): ‘Weight and water loss in the neonate in natural and forced convection’,Arch. Disease Childhood,59, pp. 951–956Google Scholar
  20. Van Doormal, J. P., andRaithby, G. D. (1984): ‘Enhancements of the SIMPLE method for predicting incompressible fluid flows’,Numerical Heat Transfer,7, pp. 147–163Google Scholar
  21. Yamaguchi, T., Taylor, T. W., Okino, H., Horio, H., andHasegawa, T. (1996): ‘Computational fluid mechanical study of the convective heat transfer in a closed space simulating an infant incubator’,Frontiers Med. Biol. Eng.,7, pp. 129–141Google Scholar

Copyright information

© IFMBE 2002

Authors and Affiliations

  1. 1.Department of Biomedical Engineering, College of Health ScienceYonsei UniversitySouth Korea
  2. 2.Research Institute for Medical Instrumentation & Rehabilitation EngineeringYonsei UniversitySouth Korea
  3. 3.Department of Mechanical EngineeringHalla UniversitySouth Korea

Personalised recommendations