Radiation dose to the lenses in CT of the paranasal sinuses

  • G. Moulin
  • C. Chagnaud
  • B. Le Brigand
  • J. M. Bartoli
  • S. Waultier
  • M. Espagnan
  • P. Châtenet
  • P. Dessi
  • G. Botti
Head and Neck Radiology

Abstract

The radiation dose to the lens during CT of the paranasal sinuses was measured in 20 patients. In 10 patients, a “standard” technique with axial and coronal sections was used. In the remaining 10 patients, overlapping axial sections for bidimensional reconstruction were obtained. Radiation dose was measured using thermoluminescent dosimeters. The mean dose was 22 mGy in patients who underwent “standard” CT and 42 mGy in those who underwent CT with overlapping sections. Dose was dependent on the total number of slices and of transorbital axial sections. Coronal sections did not statistically increase the radiation dose to the lenses. Although these doses were significantly below the critical dose for cataracts, CT with overlapping sections gives a considerably higher radiation dose. We recommend the bidimensional reconstruction technique only for selected cases (e. g. preoperative assessment of the roof of the ethmoid sinuses or in patients who can or should not maintain the hyperextended position required for coronal views).

Key words

Lens, ocular Sinuses, paranasal Computed tomography Radiation, diagnostic procedures 

References

  1. 1.
    Teatini G, Simotti G, Salvolini U, Masala W, Meloni F, Kovasio S, Dedola GL (1987) Computed tomography of the ethmoid labyrinth and adjacent structures. Ann Otol Rhinol Laryngol 96: 239–250Google Scholar
  2. 2.
    Siddle KJ, Sim LH, Case CC (1990) Radiation dose to the lens of the eye during computerised tomography of the orbit: a comparison of four modern computerised tomography units. Australas Radiol 34: 323–325Google Scholar
  3. 3.
    Neufang KFR, Zanella FE, Ewen K (1987) Radiation doses to the eye lenses in computed tomography of the orbit and the petrous bone. Eur J Radiol 7: 203–205Google Scholar
  4. 4.
    Möller I, Kühne D, Vogel H, Borchers HD, Löhr: H (1982) Strahlenex Position and Strahlenrisiko der kranialen Computertomographie. CT Sonogr 2: 1–6Google Scholar
  5. 5.
    Lawson W (1991) The intra-nasal ethmoidectomy. An experience with 1,077 procedures. Laryngoscope 101: 367–371Google Scholar
  6. 6.
    Dessi P, Moulin G, Castro F, Chagnaud C, Cannoni M (1994) Protrusion of the optic nerve in the ethmoid and sphenoid sinus. Prospective study of 150 CT examinations. Neuroradiology 36: 515–516Google Scholar
  7. 7.
    Dessi P, Moulin G, Triglia JM, Zanaret M, Cannoni M (1994) Difference in the height of the right and left ethmoid roofs: a possible risk factor for ethmoid surgery. Prospective study of 150 CT-scans. J Larnyngol Otol 108: 262–263Google Scholar
  8. 8.
    Committee on the Biological Effects of Ionizing Radiation (1990) Health effects of exposure to low levels of ionizing radiation. National Academy, Washington DC, p 363Google Scholar
  9. 9.
    Dalrymphe GV, Goulden ME, Kollmorgen GM, Vogel H (1973) Medical radiation biology. Saunders, Philadelphia, p 235Google Scholar
  10. 10.
    Lund E, Halaburt H (1982) Irradiation dose to the lens of the eye during CT of the head. Neuroradiology 22: 181–184Google Scholar
  11. 11.
    Yeoman LJ, Howarth L, Britten A, Adam EJ (1992) Gantry angulation in brain CT: dosage, implications, effect on posterior fossa artifacts, and current international practice. Radiology 184: 113–116Google Scholar
  12. 12.
    Siddle KJ, Sim L (1990) Radiation dose to the lens of the eye during computerised tomography examinations of the orbit, the pituitary fossa and the brain on a General Electric 9800 Quick CT scanner. Australas Radiol 34: 326–330Google Scholar
  13. 13.
    Torizuka T, Hayakawa K, Satoh Y, Tanaka F, Saitoh H, Okuno Y, Ogura A, Nakayama Y, Konishi J (1992) High-resolution CT of the temporal bone: a modified baseline. Radiology 184: 109–1llGoogle Scholar
  14. 14.
    McCullough EC, Payne JT (1978) Patient dosage in computed tomography. Radiology 129: 457–463Google Scholar
  15. 15.
    Shaffer KA, Littelton JT, Durizch, Callahan WP (1982) Temporal bone anatomy: comparison of computed tomography and complex motion tomography. Head Neck Surg 4: 296–300Google Scholar
  16. 16.
    Moström U, Ytterbergh C, Bergström K (1986) Eye lens dose in cranial computed tomography with reference to the technical development of the CT scanners. Acta Radiol Diagn 27: 599–606Google Scholar

Copyright information

© Springer-Verlag 1996

Authors and Affiliations

  • G. Moulin
    • 1
  • C. Chagnaud
    • 1
  • B. Le Brigand
    • 1
  • J. M. Bartoli
    • 1
  • S. Waultier
    • 2
  • M. Espagnan
    • 3
  • P. Châtenet
    • 3
  • P. Dessi
    • 4
  • G. Botti
    • 5
  1. 1.Department of RadiologyCHU la TimoneMarseille Cedex 5France
  2. 2.Department of RadiotherapyCHU la TimoneMarseilleFrance
  3. 3.Dosimetry LaboratoryCOGEMA, Etablissement de Marcoule BP 170Bagnols-sur-Ceze CedexFrance
  4. 4.ENT DepartmentCHU la TimoneMarseilleFrance
  5. 5.Department of StatisticsCHU la TimoneMarseilleFrance

Personalised recommendations