Summary
The analysis of polychlorodibenzodioxins and polychlorodibenzofurans on a novel multidimensional (CGC-CGC-MS) configuration is presented. The system is based on flow controlled column switching, allowing the simultaneous isomeric separation of tetrachloro-, pentachloro- and hexachloro-substituted dibenzodioxins and dibenzofurans on a polar column and the analysis of heptachloro- and octachloro-substituted dibenzodioxins and dibenzofurans on an apolar column. PTV injection in the solvent vent mode allows analysis of large sample sizes approaching in this way the sensitivity of CGC—high resolution MS.
Similar content being viewed by others
References
K. Ballschmiter, R. Backer, A. Mennel, R. Fisher, U. Riehle, M. Swerev, HRC15, 260 (1992).
G. Schomburg, in: Sample Introduction in Capillary Gas Chromatography, Vol 1,P. Sandra (Ed.), Hüthig Verlag, Heidelberg, 1985, p. 235.
G. Schomburg, H. Husmann, E. Hübinger, HRC & CC9, 395 (1985).
K. Ballschmiter, H. Buchert, R. Niemczyk, A. Munder, M. Swerev, Chemosphere15 901 (1986).
Author information
Authors and Affiliations
Additional information
Dedicated to Professor Leslie S. Ettre on the occasion of his 70th birthday.
Rights and permissions
About this article
Cite this article
David, F., Sandra, P., Hoffmann, A. et al. A novel CGC-CGC-MS configuration for the analysis of polychlorodibenzodioxins and polychlorodibenzofurans. Chromatographia 34, 259–265 (1992). https://doi.org/10.1007/BF02268355
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02268355