Effect of local flow disturbance on readings of electromagnetic flowmeters

  • Mechanical Measurements
  • Published:
Measurement Techniques Aims and scope

Conclusion

The output signal of a flowmeter is distorted on a pipe length of the measuring section L≤22d (Figs. 2 and 3), when the source of disturbance is located before the flowmeter and on length L≤11d when the source is located beyond the flowmeter.