Summary
A high performance liquid chromatographic method for the determination of flecainide in serum has been developed. The analysis is performed on a microparticulate silica column. The eluate is monitored by fluorescence detection at an excitation wavelength of 300nm and an emission wavelength of 370nm. No sources of interference were identified and a coefficient of variation of less than 8% was observed on repeated flecainide determinations. The method has a good reproducibility, specificity and accuracy, and can be applied in therapeutic drug monitoring of flecainide in patients.
Similar content being viewed by others
References
R. S. Bexton, K. J. Hellestrand, A. W. Nathan, R. A. J. Spurrell, A. J. Camm, Therapiewoche,34, 4926 (1984).
K. P. Kunze, M. Keutsch, K. H. Kuck, Z. Kardiol.76, (Suppl. 1), 48 (1987).
J. C. Anderson, J. R. Stewart, N. Engl. J. Med.305, 473 (1981).
D. E. Ward, M. Cheesman, M. Dancy, Int. J. Cardiol.10, 1 (1985).
G. J. Conard, G. L. Carlson, J. W. Frost, R. E. Ober, Clin. Pharmacol. Therap.25, 218 (1979).
G. J. Conard, G. L. Carlson, J. W. Frost, R. E. Ober, Clin. Therapeutics6, 643 (1984).
J. D. Johnson, G. L. Carlson, J. M. Fox, A. M. Miller, S. F. Chang, G. J. Conard, J. Pharm. Sci.73, 1469 (1984).
S. F. Chang, A. M. Miller, M. J. Jernberg, R. E. Ober, G. J. Conard, Arzneim.-Forsch./Drug Res.33 (1), 251 (1983).
J. W. DeJong, J. A. J. Hegge, E. Harmsen, P. Ph. De Tombe, J. Chromatogr.229, 498 (1982).
R. K. Bhamra, R. J. Flanagan, D. W. Holt, J. Chromatogr.307, 439 (1984).
S. F. Chang, T. M. Welscher, A. M. Miller, R. E. Ober, J. Chromatogr.272, 341 (1983).
S. F. Chang, A. M. Miller, J. M. Fox, T. M. Welscher, Therapeutic Drug Monitoring6, 105 (1984).
J. U. Becker, J. Clin. Chem. Clin. Biochem.22, 389 (1984).
F. Susanto, S. Humfeld, A. Neumann, Fresenlus Z. Anal. Chem.321, 177 (1985).
F. Susanto, S. Humfeld, H. Reinauer, Chromatographia Vol.22, 135 (1986).