Summary
Effects of column temperature on the retention behaviour of aromatic hydrocarbons and dialkyl phthalates were investigated in capillary supercritical fluid chromatography (SFC) with carbon dioxide as the mobile phase. Negative temperature programming could partly replace pressure programming. Positive temperature programming was applicable to solutes with proper volatility, in which gas chromatography-like retention mechanism (partition process) was involved.
Similar content being viewed by others
References
M. Novotny, W. Bertsch, A. Zlatkis, J. Chromatogr.,61, 17 (1971).
J. E. Conaway, J. A. Graham, L. B. Rogers, J. Chromatogr. Sci.,16, 102 (1978).
T. Takeuchi, Y. Hashimoto, D. Ishii, J. Chromatogr.,402, 328 (1987).
S. M. Fields, M. L. Lee, J. Chromatogr.349, 305 (1985).
C. R. Yonker, B. W. Wright, R. C. Pertersen, R. D. Smith, J. Phys. Chem.,89, 5526 (1985).
T. L. Chester, D. P. Innis, in P. Sandra, W. Bertsch (Ed.), Proc. Sixth Int. Symp. Capillary Chromatography, Riva del Garda, May 14–16, 1985, Rüthig, Heidelberg, 1985, p. 944.
M. L. Lee, K. E. Markides, in D. Ishii, K. Jinno, P. Sandra (Ed.), Proc. Seventh Int. Symp. Capillary Chromatography, Gifu, May 11–14, 1986, the University of Nagoya Press, Nagoya, 1986, p. 575.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Takeuchi, T., Ohta, K. & Ishii, D. Temperature programming elution in capillary supercritical fluid chromatography. Chromatographia 25, 125–128 (1988). https://doi.org/10.1007/BF02259028
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02259028