Abstract
A technique for isotopic submicrosample analysis of uranium by TIMS is presented. The heating pattern, used was modified in order to improve the precision for microsample analysis of SRM U-500 /NBS/ The technique was applied to submicrosample isotopic analysis of SRM U-150 /NBS/ and found satisfactory.
Similar content being viewed by others
References
L.P. Dunstan, J.W. Gramlich, I.L. Barnes,J. Res. Natl. Bur. Stand., 85 /1980/ 1.
R.F. Cretella, R.A.: Lukaszew, J.G. Marrero, R. Servant,Int. J. Mass Spectrom. and Ion Proc., /1989/ /to be published/.
H. Bokelund, Isotopic Analysis of Submicrogram Amounts of Uranium, Rep. ETR-199, Eurochemie, Belgium, 1966.
R.J. Jones, /Ed./ Selected Measurement Methods for Plutonium and Uranium in the Nuclear Fuel Cycle, TID 7029, Method 2500, USAEC, Washington, 1966.
W.R. Shields, The Evolution of the Accuracy of Isotopic Analysis by Thermal Ionization from 2% to 0.2%. Paper SM-67/41 from IAEA Symposium on Nuclear Materials Management, Vienna, 1965.
W.R. Shields, /Ed./ Analytical Mass Spectrometry Section: Instrumentation and Procedures for Isotopic Analysis, NBS Tech. Note 277, Washington, 1966.
Author information
Authors and Affiliations
Additional information
Thermal Ionization Mass Spectrometry.
Rights and permissions
About this article
Cite this article
Cretella, R.F., Lukaszew, R.A., Dal Fávero, J. et al. Uranium submicrosample isotopic analysis by TIMS. Journal of Radioanalytical and Nuclear Chemistry Letters 144, 379–386 (1990). https://doi.org/10.1007/BF02218146
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02218146