Skip to main content
Log in

An automated micro-method for the enzymatic determination of D-B-hydroxybutyrate in ruminant plasma

  • Research Articles
  • Published:
Veterinary Research Communications Aims and scope Submit manuscript

Abstract

  1. 1.

    An enzymatic ‘reaction rate’ micro-method for the rapid routine estimation of D-B-hydroxybutyrate (D-B-HOB) in ruminant plasma, using an I.L. Multistat III centrifugal analyzer, is described.

  2. 2.

    Reaction conditions were optimized to give a linear response for plasma D-B-HOB concentrations between 100 and 2500 μmoles per litre, at 30°C and pH 9.0.

  3. 3.

    For the standardized method the within-run and between-run coefficients of variation for deproteinised ovine plasma were consistently less than 3.5%.

  4. 4.

    There was good agreement between plasma concentrations obtained by the present method and both original U.V. end-point technique (r=0.927b=0.950) and a colorimetric end-point procedure (r=0.937. b=0.879).

  5. 5.

    Utreated ovine and bovine plasma consistently exhibited high ‘blank’ activity and this was directly correlated with plasma lactate dehydrogenase (LDH) activity in both species (r=0.971; p<0.001 and r=0.949; p<0.001 respectively). The distribution of LDH activity in man was similar to sheep but, contrastingly, non-specific interference was extremely low in human plasma and unrelated to LDH. Horse, chicken and rat had negligible ‘blank” activity and comparatively low LDH levels. In both cattle and sheep non-specific interference was abolished by perchloric acid precipitation. In the sheep subtraction of ‘blank’ activity gave D-H-HOB concentrations for untreated plasma comparable to those in deproteinised samples. However, in the bovine, D-B-HOB levels remained significantly (t=6.44; p<0.001) higher even after ‘blank’ correlation. In contrast to man and other non-ruminants, perchloric acid precipitation is essential in ruminants to avoid false overestimation of plasma D-B-HOB levels.

  6. 6.

    Plasma with EDTA as anticoagulant and serum gave concentrations of D-B-HOB approximately 60% lower, than samples containing heparin or oxalate/fluoride. However, heparin was associated with much higher (up to 50%) non-specific NAD rduction than oxalate/fluoride.

  7. 7.

    High levels of acetoacetate (400–1000 μmoles per litre) reduced the recovery of D-B-HOB from ovine plasma by less than 10%. This effect was negated by the inclusion of hydrazine hydrates in the reaction mixture. Perchlorate ion concentrations above 25 μmoles per litre per test dramatically inhibited the assay in ovine plasma, and therefore precipitation conditions must be carefully controlled.

  8. 8.

    Plasma with oxalate/fluoride as anticoagulant showed the greatest stability in storage; 24 hours at room temperature, one week at +4°C and at least one month at −20°C.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Chandrasekaran, A., Glover, R.J. and Lord, P.S. 1972. A spectrophotometric micro-method for the enzymatic determination of beta-hydroxybutyrate. Clinica Chimica Acta, 37: 371–380.

    Google Scholar 

  • Delafield, F.P. and Doudoroff, M., 1969. B-hydroxybutyrate dehydrogenase fromPseudomonas lemoigenei. Methods in Enzymology, 14: 227–231.

    Google Scholar 

  • Hansen, J.L. and Freier, E.F., 1978. Direct assays of lactate, pyruvate, B-hydroxybutyrate and acetoacetate with a centrifugal analyser. Clinical Chemistry, 24: 475–479.

    Google Scholar 

  • Kronfeld, D.D., 1972. Ketosis in pregnant sheep and lactating cows. A review. Australian Veterinary Journal, 48: 680–687.

    Google Scholar 

  • Leech, F.B., Vesey, M.P. and Macrae, W.D. 1964. Disease wastage and husbandry in the British dairy herd. Report of a national survey in 1958–59. Animal Disease Surveys, 3: 1.

    Google Scholar 

  • Li, P.K., Lee, J.T., MacGillivray, M.A., 1980. Direct fixed time kinetic assays for B-hydroxybutyrate and acetoacetate with a centrifugal analyzer or a computer-based spectrophotometer. Clinical Chemistry, 26: 1713–1717.

    Google Scholar 

  • Moore, J.J., Marcus, M. and Sax, S.M., 1982. Kinetic assay of B-hydroxybutyrate in plasma with a Cobas-Bio centrifugal analyser. Clinical Chemistry 28: 702–703.

    Google Scholar 

  • Payne, J.M., 1977. in ‘Metabolic Diseases in Farm Animals’, Heinemann, London, pp 156–166.

    Google Scholar 

  • Reid, R.L., 1968. The physiopathology of undernourishment in pregnant sheep with particular reference to pregnancy toxaemia. Advances in Veterinary Science, 12: 163–238.

    Google Scholar 

  • Williamson, D.H., Mellanby, J. and Krebs, H.W., 1962. Enzymic determination of D(-)-B-hydroxybutric acid and acetoacetic acid in blood. Biochemical Journal, 82: 90–96.

    Google Scholar 

  • Zivin, J.A. and Snarr, J.F., 1973. An automated colorimetric method for the measuremenet of 3-hydroxybutyrate concentration. Analytical Biochemistry, 52: 456–461.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Knox, D.P., Jones, D.G. An automated micro-method for the enzymatic determination of D-B-hydroxybutyrate in ruminant plasma. Vet Res Commun 8, 103–115 (1984). https://doi.org/10.1007/BF02214702

Download citation

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02214702

Keywords

Navigation