Journal of Clinical Monitoring

, Volume 12, Issue 6, pp 421–428 | Cite as

Reflectance pulse oximetry at the forehead of newborns: The influence of varying pressure on the probe

  • A. Carin M. Dassel
  • Reindert Graaff
  • Anneke Meijer
  • Willem G. Zijlstra
  • Jan G. Aarnoudse
Original Articles


Objective: Transmission pulse oximetry (TPO) is not a practical method of intrapartum fetal monitoring of arterial oxygen saturation. Reflectance pulse oximetry (RPO) requires a sensor applied to the skin of the fetal head and may be a useful technique. During labor, various degrees of pressure will be exerted on the RPO sensor. Previous studies have shown that moderate pressure on the sensor can improve the RPO signal. At increasing pressure, however, blood flow underneath the sensor will be occluded. This study examines the influence of pressure applied to the RPO sensor on the signal from the forehead of healthy newborns as a model for the fetal situation.Methods: After institutional approval, 12 healthy newborns were studied. The RPO probe was placed at the forehead. Pressure on the probe was increased stepwise from 0 to 80 mmHg, and the effect on the ratio between the relative changes of the red and infrared light intensities (R/IR, inversely related to oxygen saturation) and pulse sizes was evaluated. Additionally, the effect of firm pressure (>150 mmHg) on the probe was evaluated.Results: R/IR values remained virtually unchanged when pressure onto the probe was increased from 0 to 80 mmHg, although the standard deviation slightly decreased. The pulse size increased as pressure on the probe increased. During firm pressure on the probe (> 150 mmHg), plethysmographic signals remained detectable, but R/IR values markedly increased.Conclusions: In newborns, mild to moderate pressure on the probe has little influence on the RPO signal at the forehead. Even during firm pressure, RPO can be used to obtain pulsatile signals, that presumably derive from tissue underneath the skull, such as the cerebral circulation.

Key words

Reflectance pulse oximetry fetal monitoring newborn arterial oxygen saturation 


Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.


  1. 1.
    Dildy GA, Clark SL, Loucks CA. Preliminary experience with intrapartum fetal pulse oximetry in humans. Obstet Gynecol 1993; 81: 630–635PubMedGoogle Scholar
  2. 2.
    Gardosi JO, Schram CM, Symonds EM. Adaptation of pulse oximetry for fetal monitoring during labour. Lancet 1991; 37: 1265–1267CrossRefGoogle Scholar
  3. 3.
    Johnson N. Development and potential of fetal pulse oximetry. Contemp Rev Obstet Gynaecol 1991; 3: 193–200Google Scholar
  4. 4.
    McNamara H, Chung DC, Lilford R, and Johnson N. Do fetal pulse oximetry readings correlate with cord blood oxygenation and acidaemia? Br J Obstet Gynaecol 1992; 99: 735–738PubMedGoogle Scholar
  5. 5.
    Severinghaus JW, Kelleher JF. Recent developments in pulse oximetry. Anesthesiology 1992; 76: 1018–1038PubMedCrossRefGoogle Scholar
  6. 6.
    Graaff R, Dassel ACM, Aarnoudse JG, Zijlstra WG, Heida P, de Mul FFM, Koelink MH, Greve J. Biophysical aspects of reflection pulse oximetry. In: Lafeber HN, Aarnoudse JG, Jongsma HW, eds. Fetal and neonatal physiological measurements. Amsterdam: Elsevier, 1991: 129–134Google Scholar
  7. 7.
    Mendelson Y, Solomita MV. The feasibility of spectrophotometric measurements of arterial oxygen saturation from the fetal scalp utilizing noninvasive skin-reflectance pulse oximetry. Biomed Instrum Technol 1992; 26: 215–224PubMedGoogle Scholar
  8. 8.
    Graaff R, Dassel ACM, Koelink MH, Aarnoudse JG, de Mul FFM, Zijlstra WG, Greve J. Condensed Monte Carlo simulations applied to reflectance pulse oximetry. Proc SPIE 1993; 1888: 201–212CrossRefGoogle Scholar
  9. 9.
    Mendelson Y, McGinn MJ. Skin reflectance pulse oximetry:In vivo measurements from the forearm and calf. J Clin Monit 1991; 7: 7–12PubMedCrossRefGoogle Scholar
  10. 10.
    Shimada Y, Nakashima K, Fujiwara Y, Komatsu T, Kawanishi M, Takezawa J, Takatani S. Evaluation of a new reflectance pulse oximeter for clinical applications. Med Biol Eng Comput 1991; 29: 557–561PubMedCrossRefGoogle Scholar
  11. 11.
    Takatani S, Davies C, Sakakibara N, Zurick A, Kraenzler E, Golding LR, Noon GP, Nose Y, DeBakey ME. Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor. J Clin Monit 1992; 8: 257–266PubMedCrossRefGoogle Scholar
  12. 12.
    Dassel ACM, Graaff R, Aarnoudse JG, Elstrodt JM, Heida P, Koelink MH, de Mul FFM, Greve J. Reflectance pulse oximetry in fetal lambs. Pediatr Res 1992; 31: 266–269PubMedCrossRefGoogle Scholar
  13. 13.
    Dassel ACM, Graaff R, Sikkema M, Meijer A, Zijlstra WG, Aarnoudse JG. Reflectance pulse oximetry at the forehead improves by pressure on the probe. J Clin Monit 1995; 11: 237–244PubMedCrossRefGoogle Scholar
  14. 14.
    König V, Ullrich GJ, Huch A, Huch R. Reflection pulse oximetry — experience in Zürich. In: Lafeber HN, Aarnoudse JG, Jongsma HW. Fetal and neonatal physiological measurements. Amsterdam: Elsevier, 1991: 111–117Google Scholar
  15. 15.
    Langton JA, Hanning CD. Effect of motion artefact on pulse oximeters evaluation of four instruments and finger probes. Br J Anaesth 1990; 65: 564–570PubMedCrossRefGoogle Scholar
  16. 16.
    Wukitsch MW, Petterson MT, Tobler DR, Pologe JA. Pulse oximetry: Analysis of theory technology, and practice. J Clin Monit 1988; 4: 290–301PubMedCrossRefGoogle Scholar
  17. 17.
    Bucher HU, Keel M, Wolf M, Siebenthal KV, Duc G. Artifactual pulse oximetry estimation in neonates. Lancet 1994; 343: 1135–1136PubMedCrossRefGoogle Scholar
  18. 18.
    Eldridge FL, Hultgren HN, Wigmore ME. The physiologic closure of the ductus arteriosus in newborn infants: A preliminary report. Science 1954; 119: 731–735PubMedCrossRefGoogle Scholar
  19. 19.
    Dimich I, Singh PP, Adell A, Hendler M, Sonnenklar N, Jhaveri M. Evaluation of oxygen saturation monitoring by pulse oximetry in neonates in the delivery system. Can J Anaesth 1991; 8: 985–988CrossRefGoogle Scholar
  20. 20.
    Meier-Stauss P, Bucher HU, Hürlimann R, König V, Huch R. Pulse oximetry used for documenting oxygen saturation and right-to-left shunting immediately after birth. Eur J Pediatr 1990; 149: 851–855PubMedCrossRefGoogle Scholar
  21. 21.
    Koch G, Wendel H. Adjustment of arterial blood gases and acid base balance in the normal newborn infant during the first week of life. Biol Neonat 1968; 12: 136–161PubMedGoogle Scholar
  22. 22.
    Graaff R, Dassel ACM, Koelink MH, de Mul FFM, Aarnoudse JG, Zijlstra WG. Optical properties of human dermisin vitro andin vivo. Appl Opt 1993; 32: 435–447Google Scholar
  23. 23.
    Giacometti L. The anatomy of the human scalp. In: Montagna W, ed. Advances in biology of skin, vol VI. Oxford: Pergamon Press, 1965: 97–120Google Scholar
  24. 24.
    Ryan TJ. Development of the cutaneous circulation. In: Bralow L, ed. Fetal and neonatal physiology. Philadelphia: WB Saunders Company, 1992: 555–565Google Scholar
  25. 25.
    Perera P, Kurban AK, Ryan TJ. The development of the cutaneous microvascular system in the newborn. Br J Dermatol 1970; 82: 86–91.CrossRefGoogle Scholar
  26. 26.
    Holbrook KA 1982 A histological comparison of infant and adult skin. In: Maibach HI, Boisits EK, eds. Neonatal skin structure and function. New York: Marcel Dekker, 1982: 24–31Google Scholar
  27. 27.
    Jöbsis FF. Noninvasive infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 1977; 198: 1264–1267PubMedCrossRefGoogle Scholar
  28. 28.
    Jöbsis-van der Vliet FF, Piantadosi CA, Sylvia AL, Lucas SK, Keizer HH. Near-infrared monitoring of cerebral oxygen sufficiency. Neurol Res 1988; 10: 7–17.Google Scholar
  29. 29.
    McCormick PW, Stewart M, Lewis G, Dujovny M, Ausman JI. Intracerebral penetration of infrared light. J Neurosurg 1992; 76: 315–318PubMedGoogle Scholar
  30. 30.
    Suichies HE, Brouwer C, Aarnoudse JG, Jentink JW, de Mul FFM, Greve J. Skin blood flow changes, measured by laser Doppler flowmetry in the first week after birth. Early Hum Dev 1990; 23: 1–8.PubMedCrossRefGoogle Scholar
  31. 31.
    Martin RJ, Okken A, Rubin D. Arterial oxygen tension during active and quiet sleep in the normal neonate. J Pediatr 1979; 942: 271–274.Google Scholar
  32. 32.
    Mok JYQ, McLaughlin FJ, Pintar M, Hak H, Amaro-Galvez R, Levison H. Transcutaneous monitoring of oxygenation: What is normal? J Pediatr 1986; 108: 165–371.Google Scholar

Copyright information

© Kluwer Academic Publishers 1996

Authors and Affiliations

  • A. Carin M. Dassel
    • 1
  • Reindert Graaff
    • 1
  • Anneke Meijer
    • 1
  • Willem G. Zijlstra
    • 1
  • Jan G. Aarnoudse
    • 1
  1. 1.Department of Obstetrics & GynaecologyUniversity Hospital GroningenGZ GroningenThe Netherlands

Personalised recommendations