Fehlermöglichkeiten bei der bodyplethysmographischen Bestimmung des intrathorakalen Gasvolumens

Errors of measuring intrathoracic gas volume by body plethysmography

  • Published:
Pneumonologie Aims and scope Submit manuscript

Abstract

Under certain conditions the plethysmographic measurement of intrathoracic gas volume may include measurement of the abdominal gas volume: If the patient makes Tiffeneau-likeexpiration efforts during shutter-curve registration pressure rises not only in the thorax but also in the abdomen. This error can be avoided by asking the patient for a sigh-likeinspiration effort. Another source of an incorrect measurement of intrathoracic gas volume is the wide-spread use of high-frequency “panting”. The breathing movements have to be as slow as possible.

Zusammenfassung

In die bodyplethysmographische Meßgröße „intrathorakales Gasvolumen“ kann das abdominale Gasvolumen mit eingehen — nämlich dann, wenn der Patient während der Verschlußdruckregistrierung eine Tiffeneau-ähnlicheExspirationsbewegung ausführt und dadurch den Druck im Brust-und Bauchraum erhöht. Der Fehler läßt sich vermeiden, indem man den Patienten zu einer Seufzer-ähnlichenInspirationsbewegung auffordert. Eine zweite Fehlerquelle ist das weit verbreitete hochfrequente „Hecheln“; die Atembewegungen sollten statt dessen möglichst langsam sein.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Bedell, G. N., Marshall, R., Dubois, A. B., Harris, J. H.: Measurement of the volume of gas in the gastrointestinal tract. Values in normal subjects and ambulatory patients. J. clin. Invest35, 336–345 (1956)

    PubMed  Google Scholar 

  2. Blair, H. A., Dern, R. J., Bates, P. L.: The measurement of volume of gas in the digestive tract. Amer. J. Physiol.149, 688–707 (1947)

    Google Scholar 

  3. Du Bois, A. B., Botelho, S. Y., Bedell, G. N., Marshall, R., Comroe, J. H.: A rapid plethysmographic method for measuring thoracic gas volume: A comparison with a nitrogen washout method for measuring functional residual capacity in normal subjects. J. clin. Invest.35, 322–326 (1956)

    PubMed  Google Scholar 

  4. Nolte, D., Reif, E., Ulmer, W. T.: Die Ganzkörperplethysmographie. Methodische Probleme und Praxis der Bestimmung des intrathorakalen Gasvolumens und der Resistance-Messung bei Spontanatmung. Respiration25, 14–34 (1968)

    PubMed  Google Scholar 

  5. Smidt, U., Muysers, K.: Kritische Betrachtungen zu den methodischen Grundlagen der Ganzkörperplethysmographie. Respiration25, 116–139 (1968)

    PubMed  Google Scholar 

  6. Stadeler, H.-J.: Reproduzierbarkeit und Fehlerbreite der Residualvolumenbestimmung mit Hilfe der Ganzkörperplethysmographie. Inaugural-Dissertation, Bonn 1967

    Google Scholar 

  7. Woitowitz, H.-J., Buchheim, F. W., Woitowitz, R.: Zur Theorie und Praxis der Ganzkörperplethysmographie in der Lungenfunktionsanalyse. Prax. Pneumol.21, 449–471 (1967)

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Berger, D., Nolte, D. Fehlermöglichkeiten bei der bodyplethysmographischen Bestimmung des intrathorakalen Gasvolumens. Pneumologie 150, 305–309 (1974). https://doi.org/10.1007/BF02179334

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02179334

Key words

Navigation