Abstract
A radioisotope-excited X-ray fluorescence analysis is applied for the determination of yttrium in xenotime ore. An annular109Cd source is employed for excitation of characteristic K X-rays of yttrium. The dilution method using pure starch as a diluent has been examined. Critical thickness of diluted ore sample has been determined. For the estimation of yttrium concentration in xenotime ore, the scattered radiation technique has been utilized. The validity and precision of the method have also been investigated.
Similar content being viewed by others
References
P. VAN ESPEN, L. VAN T'DACK, F. ADAMS, R. E. VAN GRIEKEN, Anal. Chem., 51 (1979) 961.
P. ARIKAN, I. ALKAN, J. Radioanal. Chem., 67 (1981) 403.
J. J. LABRECQUE, J. M. BENSEN, R. E. VAN GRIEKEN, X-Ray Spectrom., 15 (1986) 13.
P. ARIKAN, A. ZARARSIZ, N. EFE, J. Radioanal. Nucl. Chem., 100 (1986) 303.
C. J. CARMAN, Develop. Appl. Spectrosc., 5 (1966) 45.
G. ANDERMANN, J. W. KEMP, Anal. Chem., 30 (1958) 1306.
M. OKADA, Y. KAMEMOTO, J. Chem., Soc. Japan, Pure Chem. Sect., 85 (1964) 641.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Leenanupan, V., Ratanalert, N. Determination of yttrium in xenotime ore by energy dispersive X-ray fluorescence spectrometry. Journal of Radioanalytical and Nuclear Chemistry, Articles 131, 385–390 (1989). https://doi.org/10.1007/BF02060604