Abstract
We describe a radioimmunoassay procedure for the measurement of phenytoin (5,5-diphenylhydantoin) in serum samples. Antiserum to phenytoin has been produced against phenytoin-valeric acid-bovine serum albumin conjugate.125I labelled phenytoin-acetic acid-tyrosine methyl ester has been used as a tracer. The assay covers a range of 10–500 ng/cm3 and has a sensitivity of 0.25 ng. The assay is validated by specificity tests, precision profile and recovery tests.
Similar content being viewed by others
References
F. BUCHTHAL, O. SEVENSMARK, AMA Archives of Neurol., 2 (1960) 624.
W. H. REYNOLDS, D. CHADWICK, A. W. GALBRAITH, Lancet, 1 (1976) 923.
F. ROCHNER, W. D. HOOPER, J. H. TYRER, M. J. EADIE, J. Neurol. Neurosurg. Psych., 35 (1972) 873.
A. RICHENS, A. DUNLOP, Lancet, 2 (1975) 247.
J. W. PAXTON, F. J. ROWELL, J. G. RATCLIFFE, J. Immunol. Methods, 10 (1976) 317.
W. M. HUNTER, F. C. GREENWOOD, Nature (London), 194, (1962) 495.
C. E. COOK, J. A. KEPLER, H. D. CHRISTENSEN, Res. Com. Chem. Pathol. Pharmacol., 5 (1973) 767.
GRACE SAMUEL, N. SIVAPRASAD, R. S. MANI, Symp. on Radiochemistry and Radiation Chemistry, Preprints, IGCAR, Kalpakkam, India, 1989.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Samuel, G., Sivaprasad, N. & Mani, R.S. Radioimmunoassay of phenytoin. Journal of Radioanalytical and Nuclear Chemistry, Articles 149, 3–11 (1991). https://doi.org/10.1007/BF02053707