Summary
The electronic determination of volume of the corpuscular blood components is more accurate and quicker than optical methods. The main source of error with the coulter instrument is the unequal current density in the cell-sizing aperture. This error was discovered by comparative examinations with the electron microscope, experiments using an enlarged model (electrolytic tank), and oscillographic puls shape analysis. This error in the measurement system causes the artificial positive skewness of the volume distribution curve. This error is avoided using a new device which permits exact volume determination in a homogeneous electric field without coincidence error. Form factors can be neglected in the determination of erythrocyte volume. They are only of importance in measuring the absolute volume of leukocytes as well as for calibrating instruments with latex-particles.
Zusammenfassung
Die elektronische Volumenbestimmung der corpusculären Bestandteile des Blutes zeichnet sich gegenüber optischen Methoden durch grö\ere Genauigkeit und eine sehr hohe Me\geschwindigkeit aus. Die wesentliche Fehlerquelle bei den Coulterschen Geräten, die unterschiedliche Stromdichte in der Me\öffnung der Glascapillare, wurde durch elektronenmikroskopische Referenzuntersuchungen, Messungen am vergrö\erten Modell und Impulsbildanalysen aufgedeckt. Dieser systematische Fehler verursacht die artifizielle Rechtsschiefe der Erythrocytenvolumenverteilungskurve. Er entfällt bei Anwendung einer modifizierten Capillare, die unverfälschte Volumenbestimmungen durch Messungen im homogenen elektrischen Feld ermöglicht. — Formfaktoren können für die Volumenbestimmung der Erythrocyten vernachlässigt werden. Ihre Berücksichtigung ist zur Ermittlung des Absolutvolumens der Leukocyten sowie zur Gerätekalibrierung mit Latexpartikeln unerlä\lich.
Similar content being viewed by others
Literatur
Adelkofer, F., Uher, L., Ketnath, H. J., Ruhenstroth-Bauer, G.: Einflu\ von Lysolecithin und unveresterten Fettsäuren auf die Blutkörperchensenkungsgeschwindigkeit. Blut19, 321 (1969).
Brecher, G., Jakobek, E. F., Schneidermann, M. A., Williams, G. Z., Schmidt, P. J.: Size distribution of erythrozytes. Ann. N.Y. Acad. Sci.99, 242 (1962).
Bull, B. S.: On the distribution of red cell volumes. Blood31, 503 (1968).
Cole, K. S., Curtis, H. J.: Electric Physiology. Md. Physics, Vol.1, ed. O. Glasser. Chicago: Year Book Publ. 1944.
Coopersmith, A., Ingram, M.: Red cell volumes and erythropoiesis I. Age density: volume relationship of normocytes. Amer. J. Physiol.215, 1276 (1968).
Coulter, W. H.: Theory of the Coulter counter, Coulter electronics. Florida: Inc. Hialeah.
Dilla, M. van, Basman, N. J., Fulwyler, M. J.: Los Alamos Scientific Laboratory Report. L. A.31, 503 (1964).
Donald, B., Nevius, M. D.: Osmotic error in electronic determination of red cell volume. Amer. J. Clin. Path.39, 38 (1963).
Fricke, H.: A mathematical treatment of the electric conductivity and capacity of disperse systems. Physiol. Rev.24, 575 (1924).
Grant, J. L., Britton, M. C., Kurtz, T. E.: Measurement of red blood cell volume with the electronic cell counter. Amer. J. clin. Path.33, 138 (1960).
Gregersen, C. A., Bryant, W., Hammerle, E. S., Usami, S., Chien, S.: Flow characteristics of human erythrozytes through polycarbonate sieves. Science157, 825 (1967).
Gregg, E. C. Steidley, D.: Electrical counting and sizing of mammilian cells in suspension. Biophys. J.5, 393 (1965).
Gutmann, J.: Elektronische Verfahren zur Ermittlung statistischer Ma\zahlen einiger medizinisch wichtiger Daten. Elektromedizin11, 62 (1966).
- Persönliche Mitteilung.
—, Hoffmann, G., Ruhenstroth-Bauer, G.: Standardization in Haematology III. New York: Karger 1966.
Jacobi, H., Hanstein, A., Hanstein, W., Künzer, W.: Erfahrungen mit einem elektronischen Erythrozytenzählgerät bei Zählungen und Volumenverteilungskurven. Klin. Wschr.45, 154 (1967).
Lushbaugh, C. C., Basmann, N. J., Glascock, B.: Electronic measurement of cellularvolumes II. Frequency distribution of erythrozyte volumes. Blood20, 241 (1962).
Nevius, D. B.: Osmotic error in electronic determinations of red cell volume. Amer. J. clin. Path.39, 38 (1963).
Ponder, E.: Red Blood Cell Dimensions: Measurement. Med. Physics, Vol.1, ed. O. Glasser. Chicago: Year Book Publ. 1944.
—: Haemolysis and Related Phenomenas. New York: Grune and Stratton 1948.
Ruhenstroth-Bauer, G., Zang, D.: Automatische Zählmethoden: Das Coultersche Partikelzählgerät. Blut6, 446 (1960).
Skalag, R., Branemark, P.-J.: Deformation of red blood cells in capillaries. Science164, 717 (1969).
Thom, R.: Quantitative Untersuchungen über die Phagozytoseleistung der Granulozyten unter Cytostatikaeinflu\. Internat. Kongr. Chemotherapie Wien 1967, Wiener med. Akademie.
- Coulter Symposium, Düsseldorf 1968.
- Coulter Symposium, Bad Nenndorf 1968.
—: Erythrozytenvolumenverteilungskurven in der hämatologischen Diagnostik. Dtsch. med. Wschr.94, 617 (1969).
Ur, A., Lushbaugh, C. C.: Some effects of electrical fields on red blood cells with remarks on electronic real cell sizing. Brit. J. Haemat.15, 527 (1968).
Weed, R. I., Bowdler, A. J.: The influence of hemoglobin conzentration on the distribution pattern of the volume of human erythrozytes. Blood29, 297 (1967).
Weiss, A. v., Kleinwächter, H.: übersicht über die theoretische Elektrotechnik, 2. Teil. Tüssen: C. F. Wintersche Verlagshandlung 1956.
Winter, H., Sheard, R. P.: The skewness of volume distribution curves of erythrocytes. Aust. J. exp. Biol. med. Sci.44, 687 (1965).
Zur Anwendung und Methodik der Mikrophotographie von Bewegungsabläufen mit Riesen-Impulslasern. (In Vorbereitung.)
Author information
Authors and Affiliations
Additional information
Herrn Professor Dr. Ruhenstroth-Bauer danke ich für sein Entgegenkommen, da\ wesentliche Teile dieser Arbeit in der Abteilung für Experimentelle Medizin, Max-Planck-Institut für Biochemie, München, durchgeführt werden konnten. Herrn Dipl.-Ing. V. Kachel wird für die Mithilfe bei elektronischen Messungen und Durchsicht des Manuskriptes, Herrn Dr. F. Adelkofer für wertvolle Anregungen zur Lysolecithin-Präparation gedankt.
Rights and permissions
About this article
Cite this article
Thom, R., Hampe, A. & Sauerbrey, G. Die elektronische Volumenbestimmung von Blutkörperchen und ihre Fehlerquellen. Z. Gesamte Exp. Med. 151, 331–349 (1969). https://doi.org/10.1007/BF02052922
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02052922