Skip to main content
Log in

Evaluation of the MS-2 automated antimicrobial susceptibility testing system: Report of a European collaborative study

  • Original Articles
  • Published:
European Journal of Clinical Microbiology Aims and scope Submit manuscript

Abstract

The MS-2 is an instrument for rapid automated testing of antimicrobial susceptibility. Its performance was evaluated by comparison with disc diffusion and MIC tests in a collaborative study in four European laboratories. Tests on independently isolated organisms showed the MS-2 to be in essential agreement with conventional methods in 94.8% of tests. A further series of 170 welldefined pathogens for which a reference antibiogram was available were tested by MS-2 and the manual methods in all of the participating laboratories. MS-2 results were in full accord in 90 % of tests and in essential agreement in 94 %. MS-2 results compared at least as well with the reference values as did either of the manual methods. Initial problems of false susceptibility results with erythromycin and penicillin were resolved by (1) the addition of small amounts of erythromycin which acted as an inducer and (2) by the use of a low content penicillin disc. MS-2 was found to be reliable and needed no attention following loading of the test cuvette cartridges. A print-out of the results was available 2–5 h after inception of the test.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. McKie, J. E., Horovor, R. J., Dooley, J. F., Evanega, G. R., Mendoza, G., Meyer, F., Moody, M., Packer, D. E., Praglin, J., Smith, H.: Autobac 1 — a 3-hour automated antimicrobial susceptibility test system. In: Heden, C. G., Illeni, T. (ed.): Automation in microbiology and immunity. Wiley, New York, 1975, p. 209–241.

    Google Scholar 

  2. Spencer, H. J., Stockert, J., Welaj, P., Wilborn, R., Price, B.: Automated antibiotic susceptibility testing with the MS-2 system. In: Johnston, H. H., Newsom, S. W. B. (ed.): Rapid methods and automation in microbiology. Learned Information, Oxford, 1977, p. 272–275.

    Google Scholar 

  3. Isenberg, H. D., Sampson-Scherer, J.: Clinical laboratory feasibility study of antibiotic susceptibility determined by the Auto Microbic System. In: Nelson, J. D., Grassi, C. (ed.): Current chemotherapy and infectious disease. American Society for Microbiology, Washington, 1980, p. 526–528.

    Google Scholar 

  4. National Committee for Clinical Laboratory Standards: Performance standards for antimicrobial disc susceptibility teste. Approved standard: ASM-2. NCCLS, Villanova, 1979.

  5. Ericsson, H., Sherris, J. C.: Antibiotic sensitivity testing. Report of an international collaborative study. Acta Pathologica et Microbiologica Scandanavica (B) 1971. Supplement 217: 12–25.

    Google Scholar 

  6. Thornsberry, C., Anhalt, J. P., Washington, J. A., McCarthy, L. R., Schoenknect, F. D., Sherris, J. C., Spencer, H. J.: Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative study. Journal of Clinical Microbiology 1980, 12: 375–390.

    PubMed  Google Scholar 

  7. Bauer, A. W., Kirby, W. M. M., Sherris, J. C., Turck, M.: Antibiotic susceptibility testing by a standardised single disc method. American Journal of Clinical Pathology 1966, 45: 462–464.

    Google Scholar 

  8. Stokes, E. J., Waterworth, P. M.: Antibiotic sensitivity tests by diffusion methods. Broadsheet 55. Association of Clinical Pathologists, London, 1972.

    Google Scholar 

  9. Matsen, J. M., Krall, B. J., Saxon, B. A.: The influence of first day susceptibility testing on antimicrobial therapy and length of patient hospitalisation. In: Johnston, H. H., Newson, S. W. B. (ed.): Rapid methods and automation in microbiology. Learned Information, Oxford, 1977, p. 60.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Johnston, H.H., Acar, J.F., Linzenmeier, G. et al. Evaluation of the MS-2 automated antimicrobial susceptibility testing system: Report of a European collaborative study. Eur. J, Clin. Microbiol. 1, 204–211 (1982). https://doi.org/10.1007/BF02019709

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02019709

Keywords

Navigation