Abstract
The MS-2 is an instrument for rapid automated testing of antimicrobial susceptibility. Its performance was evaluated by comparison with disc diffusion and MIC tests in a collaborative study in four European laboratories. Tests on independently isolated organisms showed the MS-2 to be in essential agreement with conventional methods in 94.8% of tests. A further series of 170 welldefined pathogens for which a reference antibiogram was available were tested by MS-2 and the manual methods in all of the participating laboratories. MS-2 results were in full accord in 90 % of tests and in essential agreement in 94 %. MS-2 results compared at least as well with the reference values as did either of the manual methods. Initial problems of false susceptibility results with erythromycin and penicillin were resolved by (1) the addition of small amounts of erythromycin which acted as an inducer and (2) by the use of a low content penicillin disc. MS-2 was found to be reliable and needed no attention following loading of the test cuvette cartridges. A print-out of the results was available 2–5 h after inception of the test.
Similar content being viewed by others
References
McKie, J. E., Horovor, R. J., Dooley, J. F., Evanega, G. R., Mendoza, G., Meyer, F., Moody, M., Packer, D. E., Praglin, J., Smith, H.: Autobac 1 — a 3-hour automated antimicrobial susceptibility test system. In: Heden, C. G., Illeni, T. (ed.): Automation in microbiology and immunity. Wiley, New York, 1975, p. 209–241.
Spencer, H. J., Stockert, J., Welaj, P., Wilborn, R., Price, B.: Automated antibiotic susceptibility testing with the MS-2 system. In: Johnston, H. H., Newsom, S. W. B. (ed.): Rapid methods and automation in microbiology. Learned Information, Oxford, 1977, p. 272–275.
Isenberg, H. D., Sampson-Scherer, J.: Clinical laboratory feasibility study of antibiotic susceptibility determined by the Auto Microbic System. In: Nelson, J. D., Grassi, C. (ed.): Current chemotherapy and infectious disease. American Society for Microbiology, Washington, 1980, p. 526–528.
National Committee for Clinical Laboratory Standards: Performance standards for antimicrobial disc susceptibility teste. Approved standard: ASM-2. NCCLS, Villanova, 1979.
Ericsson, H., Sherris, J. C.: Antibiotic sensitivity testing. Report of an international collaborative study. Acta Pathologica et Microbiologica Scandanavica (B) 1971. Supplement 217: 12–25.
Thornsberry, C., Anhalt, J. P., Washington, J. A., McCarthy, L. R., Schoenknect, F. D., Sherris, J. C., Spencer, H. J.: Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative study. Journal of Clinical Microbiology 1980, 12: 375–390.
Bauer, A. W., Kirby, W. M. M., Sherris, J. C., Turck, M.: Antibiotic susceptibility testing by a standardised single disc method. American Journal of Clinical Pathology 1966, 45: 462–464.
Stokes, E. J., Waterworth, P. M.: Antibiotic sensitivity tests by diffusion methods. Broadsheet 55. Association of Clinical Pathologists, London, 1972.
Matsen, J. M., Krall, B. J., Saxon, B. A.: The influence of first day susceptibility testing on antimicrobial therapy and length of patient hospitalisation. In: Johnston, H. H., Newson, S. W. B. (ed.): Rapid methods and automation in microbiology. Learned Information, Oxford, 1977, p. 60.