Measurement of flow velocity in the model circulation by videodensitometry. Methodological investigations

Methodische Untersuchungen zur videodensitometrischen Messung von Flußgeschwindigkeiten im Modellkreislauf

  • Original Contributions
  • Published:
Basic Research in Cardiology Aims and scope Submit manuscript

Summary

The relation between videodensitometrically measured front velocity and electromagnetically assessed flow was examined in a circulatory model with continuous as well as pulsatile flow (89 experiments). The diameter of the tubes in the videodensitometric measuring section was 0.305 to 0.518 cm. A linear correlation was proved in flow velocities up to Reynold's number Re=225. The exact flow, measured electromagnetically, was overestimated in continuous flow by 21% (r=0.99, Syx=±14.5 ml/min) and in pulsatile flow by 24% (r=0.98, Syx=±20.8 ml/min). In view of these results the phasic and average flow can be calculated acurately using videodensitometric techniques.

Zusammenfassung

Am Modellkreislauf wurde in 89 Einzelmessungen bei kontinuierlicher und pulsierender Strömung die Beziehung zwischen videodensitometrisch mittels Frontgeschwindigkeit bestimmtem und elektromagnetisch ermitteltem Fluß untersucht. Der Rohrdurchmesser der videodensitometrischen Meßstrecke lag zwischen 0,305 cm und 0,518 cm. Zwischen den Flußgeschwindigkeiten bestand bis zu einer Reynold-Zahl von Re=225 ein linearer Zusammenhang. Dabei wurde der wahre (elektromagnetische Messung) Fluß bei kontinuierlicher Strömung um 21% (r=0,99; Syx=±14.5 ml/min) und bei pulsierender Strömung um 24% (r=0,98; Syx=±20.8 ml/min) überschätzt. Danach ist unter Berücksichtigung der Überschätzung eine videodensitometrische Bestimmung des phasischen und mittleren Flusses zuverlässig möglich.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Decker, D., F. J. Roth, E. Epple, J.-R. Allenberg: Meßabweichungen der Röntgen-Cine-Densitometrie, überprüft an Modellkreisläufen. Biomed. Tech. (Stuttg.)18, 133 (1973).

    Google Scholar 

  2. Heintzen, P. H. (Ed.): Roentgen-, Cine-and Videodensitometry (Stuttgart, 1971).

  3. Rutishauser, W.: Kreislaufanalyse mittels Röntgendensitometrie (Bern 1969).

  4. Smith, H. C., R. L. Frye, D. E. Donald, G. D. Davis, J. R. Pluth, R. E. Sturm, E. H. Wood: Roentgen Videodensitometric Measure of Coronary Blood Flow. Mayo Clin. Proc.46, 800 (1971).

    PubMed  Google Scholar 

  5. Vanselow, K., F. Heuck, H. K. Deininger: Neue Grundlagen und Theorien zur Verbesserung der Angio-Cine-Densitometrie. Fortschr.Geb. Röntgenstr. Nuklearmed.123, 5 (1975).

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

With 9 figures and 3 tables

Supported by Deutsche Forschungsgemeinschaft SFB 30

Rights and permissions

Reprints and permissions

About this article

Cite this article

Fermor, U., Huber, H., Neuhaus, K.L. et al. Measurement of flow velocity in the model circulation by videodensitometry. Methodological investigations. Basic Res Cardiol 74, 361–377 (1979). https://doi.org/10.1007/BF01908390

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01908390

Keywords

Navigation