Ventrikeldruckmessung mit konventionellen Kathetermanometersystemen

Teil I Registriertechnische Minimalforderungen an Blutdruckmeßsysteme und Bestimmung ihrer Schwingungseigenschaften

Recording of ventricular pressure by conventional catheter manometer systems

Minimal requirements of blood pressure recording systems and estimation of frequency response characteristics

  • Originalarbeiten
  • Published:
Basic Research in Cardiology Aims and scope Submit manuscript

Summary

Parallel recordings of pressure pulses by conventional catheter manometer systems and catheter tip manometer demonstrate severe errors in peak velocity of pressure rise estimated by conventional systems. This fact is due to inadequate dynamic response characteristics of conventional systems in relation to the frequency content of pressure curves. During cardiac rest the error in dp/dtmax is less than 10% if the frequency response of the recording system is uniform up to 10 Hz, the corresponding value under maximal cardiac stimulation is about 40 Hz. This is equal to the first 10 harmonics of heart rate. The examination of left and right ventricular pressure curves leads to similar results. The experimental determination of dynamic response characteristics of catheter manometer systems requires a test system producing suitable sinus or step functions, parallel high fidelity recording of pressure functions is to be recommended. A simple test station is described. Examinations of temperature influence on catheter material and resulting changes in dynamic response characteristics were carried out. The incubation of catheters at the temperature of 37° C is indispensable. A new diagram for simplifying the interpretation of results is described. Other publications are discussed in viewpoint of employed techniques and representation of results.

Zusammenfassung

Parallele Erfassungen von Druckkurven über verschiedene konventionelle Kathetermanometersysteme und Kathetertip-Manometer zeigen hohe Fehler der mit konventionellen Systemen ermittelten Druckanstiegsgeschwindigkeiten. Ursache ist die mangelnde Anpassung der Registriereigenschaften dieser Systeme an das Frequenzspektrum der Druckkurven als Meßgrößen. Druckaufnehmer ermöglichen eine Erfassung der Druckanstiegsgeschwindigkeit mit Fehlern unter 10% in Ruhe bzw. unter Belastung, wenn sie eine fehlerfreie Wiedergabe von Sinusschwingungen bis zur oberen Grenze von ca. 10 Hz bzw. 40 Hz gewährleisten. Diese Werte entsprechen jeweils etwa der 10 harmonischen Oberschwingung, bezogen auf die Herzfrequenz als Grundfrequenz, sie gelten für Druckkurven aus rechtem und linkem Ventrikel. Zur ermittlung der Schwingungseigenschaften von Kathetermanometersystemen sind Testapparaturen mit hinreichend genauen Sinus-bzw. Sprungfunktionen erforderlich, die Einbringung von Kathetertip-Manometern als Kontroll-und Bezugssystem ist zu empfehlen. Ein einfaches Testsystem wird beschrieben. Der Temperatureinfluß auf das Herzkathetermaterial und resultierende Änderungen des Schwingungsverhaltens werden untersucht, eine Inkubation der Katheter bei 37° C ist für eine praxisbezogene Beurteilung der Systeme unerläßlich. Ein neues Diagramm zur vereinfachten Interpretation der erhobenen Meßdaten wird beschrieben. Untersuchungsberichte anderer Arbeitsgruppen werden nach den Kriterien der eingesetzten Methodik und der Darstellung der Ergebnisse diskutiert.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  • Both, A., U. Gleichmann, F. Loogen, W. Mäurer, J. Ressel, Z. Kreislaufforschg.58, 1212–1224 (1970).

    Google Scholar 

  • Bretschneider, H. J., L. A. Cott, G. Hellige, I. Hensel, D. Kettler, J. Martel, A new hemodynamic parameter consisting of 5 additive determinants for estimation of the O2-consumption of the left ventricle, p. 663. Proc. internat. Union Physiol. Sci. IX. XXV. Internat. Congr. (Munich 1971).

  • Burton, A. C., Physiologie und Biophysik des Kreislaufs (Stuttgart-New York (1969).

  • Degenring, F. H., H. A. Dittmar, Z. Kardiol.62, 753–759 (1973).

    PubMed  Google Scholar 

  • Dittmar, H. A., F. H. Degenring, Z. Kreislaufforschg.60, 212–220 (1971).

    Google Scholar 

  • Fa.Edwards Laboratories, Rev. 10/72 II P 5M 2-1-430 (Santa, Ana, California, USA 1972 a).

  • Fa.Edwards Laboratories, Original Swan-Ganz Einschwemm-Rechtsherz-Katheter (München, Nov. 1972 b).

  • Fitzpatrich, F., G. Hampson, H. Burgess, Canad. med. Ass. J.106, 24–29 (1972).

    Google Scholar 

  • Forrester, S., W. Ganz, G. Diamond, T. McHugh, D. W. Chonette, H. J. C. Swan, Amer. Heart J.83, 306–311 (1972).

    Article  PubMed  Google Scholar 

  • Geddes,, L. A., The direct and indirect measurement of blood pressure (Chicago 1970).

  • Hansen, A. T., E. Warburg, Acta physiol. scand.19, 306–332 (1950a).

    Google Scholar 

  • Hansen, A. T., Acta physiol. scand19, 306–332 (1950b).

    Google Scholar 

  • Hansen, A. T., Acta physiol. scand.19, 1–228 (1950c).

    Google Scholar 

  • Hellige, G., Frequenzspektrum von Herzdruckkurven; Meßfehler durch das Resonanzverhalten konventioneller Kathetermanometersysteme und Bandbreiteneinengung. Deutsche Physiologische Gesellschaft, Referate der 40. Tagung, R 30-59 (1972).

  • Hellige, G., Grenzen der Erfassung von dp/dtmax und Ventrikeldrucken mit verschiedenen Kombinationen konventioneller Katheter, und neuer Druckaufnehmer. Habilitationsschrift (Göttingen 1973).

  • Fa.Fritz Hellige & Co. GmbH, Aufnehmer für physiologische Meßgröße. P/Meß/6909 (1969a).

  • Fa.Fritz Hellige & Co. GmbH, Elektrische Blutdruckmessung und Einschwemm-Katheterisierung. P/E-Kath + Ma 83/6911 (1969 b).

  • Knopp, J., S. H. Rahimtoola, H. J. C. Swan, Suppl. II to Circulation31, and32 128 (1965).

    Google Scholar 

  • Lilly, J. C., V. Legallis, R. Cherry, J. Appl. Physics18, 613–628 (1947).

    Article  Google Scholar 

  • Mason, D. T., R. Zelis, E. A., Amsterdam, Triangel9, 273–281 (1971).

    Google Scholar 

  • Melville, A. W., J. Sci. Instrum.36, 422–424 (1959).

    Article  Google Scholar 

  • Noble, F., IRE Trans. on Med. Electronics PGME8, 38–45 (1957).

    Google Scholar 

  • Patel, J. D., D. T. Mason, J. Ross, E. Braunwald Amer. Heart J.69, 785–799 (1965).

    Article  PubMed  Google Scholar 

  • Porje, J. G., Acta physiol. scand.42, 7–68 (1946).

    Google Scholar 

  • Roskamm, H., H. Weidemann, B. Meinecke J. Petersen, H. Reindell, Z. Kreislaufforschg.59, 119–138 (1970).

    Google Scholar 

  • Rushmer, R. F., Cardiovascular-Dynamics (Philadelphia-London-toronto 1970).

  • Fa. Dr. med.J. Schubart, Dipl-Phys.G. Schubart Übersicht über Blutdruckaufnehmer für medizinische Anwendung (Hofheim/Taunus 1966).

  • Shapiro, G. G., L. J. Kroretz, Amer. Heart J.80, 226–236 (1970).

    Article  PubMed  Google Scholar 

  • Stegall H. F., J. appl. Physiol.22, 591–592 (1967).

    PubMed  Google Scholar 

  • Sutter, W., H. J. C. Swan E. H. Wood, Physiologist1, 81 (1957).

    Google Scholar 

  • Wagner, K. W., Einführung in die Lehre von den Schwingugen und Wellen (Wiesbaden 1947).

  • Warburg, E., Acta physiol. scand.,19, 344–349 (1950).

    Google Scholar 

  • Weller, W., W. T. Ulmer, W. Walkenhorst, Z. Kreislaufforschg.58, 689–698 (1969).

    Google Scholar 

  • Wood, E. H., Science112, 707–715 (1950).

    PubMed  Google Scholar 

  • Wood, E. H., Communication and Electronics23, 32–40 (1956).

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Mit 8 Abbildungen

Mit Unterstützung der Deutschen, Forschungsgemeinschaft im Rahmen des SFB 89 Kardiologie, Göttingen.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hellige, G. Ventrikeldruckmessung mit konventionellen Kathetermanometersystemen. Basic Res Cardiol 71, 319–336 (1976). https://doi.org/10.1007/BF01906457

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01906457

Navigation