Pulse contour cardiac output as a clinically valuable tool for intensive patient monitoring. A critique of a recent paper

  • K. H. Wesseling
  • P. D. Verdouw
  • J. Beaune
  • J. Roelandt
  • P. G. Hugenholtz
Letter to the Editor

Summary

Positive as well as negative reports on pulse contour stroke volume or cardiac output measuring have appeared in vast quantity in the literature. A recent paper byVerdouw et al. in this journal is of the negative variety, but on insufficient grounds, since:
  1. 1.

    young healthy swine no not constitute a proper model for old age human cardiac patients

     
  2. 2.

    statistics and regression analysis, on which much of the negative conclusions are based, are used improperly

     
  3. 3.

    the electromagnetic flow probe cannot be used uncritically as the reference method, and

     
  4. 4.

    over 2000 comparisons, in three studies in man, of a new pulse contour method intended for intensive patient monitoring, lead to the conclusion that at least that method is reliable and has a clinically useful accuracy.

     

Keywords

Cardiac Output Stroke Volume Cardiac Patient Flow Probe Negative Variety 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Pulskontur-Herzminutenvolumen als eine klinisch wertvolle Angabe für intensive klinische Patientenüberwachung. Eine Kritik auf einen neulich erschienenen Artikel

Zusammenfassung

Positive wie auch negative Berichte über Pulskontur-Herzschlagvolumen oder Überwachung des Herzminutenvolumens sind häufig in der Literatur erschienen. Ein in dieser Zeitschrift vor kurzem erschiener Bericht vonVerdouw et al. gehört zu den negativen Variationen, jedoch ungenügend begründet, weil:
  1. 1.

    junge, gesunde Schweine kein geeignetes Modell für ältere menschliche Herzpatienten bilden,

     
  2. 2.

    Statistiken und Regressionsanalysen, auf denen viele der negativen Schlüsse beruhen, nicht richtig verwendet worden sind,

     
  3. 3.

    der elektromagnetische Strömungsaufnehmer nicht kritiklos als Referenz verwendet werden kann und

     
  4. 4.

    mehr als 2000 Vergleiche (in drei Studien an Menschen) einer neuen Pulskonturmethode für intensive Patientüberwachung zu der Schlußfolgerung führen, daß die Methode wenigstens zuverlässig ist und eine klinisch brauchbare Genauigkeit hat.

     

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Erens, H. M. C., In vitro metingen van het visco-clastisch gedrag van de aorta thoracalis descendens van de mens. Thesis Technological University Eindhoven (1975).Google Scholar
  2. 2.
    Goedhard, W. J. A., Visco-elastische eigenschappen van de arteriewand. Thesis Free University, Amsterdam (1971).Google Scholar
  3. 3.
    Goedhard, W. J. A., A. A. Knoop, A model of the arterial wall. J. Biomech.6, 281–288 (1973).PubMedGoogle Scholar
  4. 4.
    Goedhard, W. J. A., A. A. Knoop, N. Westerhof, The influence of vascular smooth muscle contraction on elastic properties of pig's thoracic aorta. In: Extrait des Acta Cardiologica28, 415–430 (1973(.PubMedGoogle Scholar
  5. 5.
    McDonald, D. A., Blood flow in arteries (2nd edition) (London 1974).Google Scholar
  6. 6.
    Nichols, W. W., Continuous cardiac output derived from the aortic pressure waveform: a review of current methods. Biomed. Engng., 376–379, sept. (1973).Google Scholar
  7. 7.
    Purschke, R., P. Brucke, H. D. Schulte, Teil II: Langzeitbeobachtung bei Patienten. Der Anästhesist23, 525 (1974).Google Scholar
  8. 8.
    Purschke, R., E. Pütz, J. O. Arndt, Untersuchung zur Zuverlässigkeit der Schlagvolumenbestimmung aus der Aortendruckkurve; Teil I: Tierexperimentelle Ergebnisse. Der Anästhesist23, 483 (1974).Google Scholar
  9. 9.
    Purschke, R., K. H. Wesseling, H. D. Schulte, Kontinuierliche Überwachung des Herzminutenvolumens mit einem neuen Recheneinschub. Medizinelektronik2, 2–7 (1975).Google Scholar
  10. 10.
    Schoenberg, A. A., U. Mennicken, R. Simon, P. Brambring, P. H. Heintzen, Stroke volume estimation from aortic pressure with correction for arrhythmias. Eur. J. Cardiol.2, 55–65 (1974).PubMedGoogle Scholar
  11. 11.
    Smith, N. Ty, K. H. Wesseting, J. A. Weber, B. de Wit, Preliminary evaluation of a pulse contour cardiac output computer in man. Feasibility of brachial or radial arterial pressures. Proc. San Diego Biomed. Symp.13, 113–119 (1974).Google Scholar
  12. 12.
    Snoeckx, L. H., J. L. Verheyen, A. van de Water, P. Lewi, R. S. Reneman, The on-line computation of cardiac output with the thermo-dilution method using a digital mini computer. Cardiovascular Research (to be published).Google Scholar
  13. 13.
    Wesseling, K. H., B. de Wit, J. A. P. Weber, Computer zur Ermittlung des Herzminutenvolumens aus der Pulskontur. Medizinische Technik70/3, 64–68 (1974).Google Scholar
  14. 14.
    Wesseling, K. H., W. W. Nichols, B. de Wit, J. A. P. Weber, A beat-to-beat cardiac output computer. Part I. Principle of operation and model evaluation, andNichols, W. W., K. H. Wesseling, B. de Wit, J. A. P. Weber, Part II. Animal experimental evaluation. Proc. 3rd ICMP, 19.5, 19.6, Göteborg (1972).Google Scholar
  15. 15.
    Wesseling, K. H., R. Purschke, N. T. Smith, H. D. Schulte, J. A. P. Weber, A beat-to-beat cardiac output computer for clinical monitoring. In: Real time computing in patient management, 92–112 (1976).Google Scholar
  16. 16.
    Wesseling, K. H., R. Purschke, N. Ty Smith, H. J. Wüst, B de Wit, J. A. P. Weber, A computer module for the continuous monitoring of cardiac output in the operating theatre and the ICU. Acta Anaesth. Belgica27, Suppl. 87–102 (1976).Google Scholar
  17. 17.
    Wesseling, K. H., N. Ty Smith, W. W. Nichols, B. de Wit, J. A. P. Weber, A small, beat-to-beat cardiac output computer. Proc. San Diego Biomed. Symp.13, 107–112 (1974).Google Scholar
  18. 18.
    Wesseling, K. H., N. Ty Smith, W. W. Nichols, J. A. P. Weber, B. de Wit, J. E. W. Beneken, Beat-to-beat cardiac output from the arterial pressure pulse contour. In:Feldman, Leigh, Spierdijk (eds.): Measurement in anesthesia. Leiden University Press, 150–164 (1974).Google Scholar

References

  1. 1.
    Wesseling, K. H., W. W. Nichols, B. de Wit, H. Weber: A beat-to-beat cardiac output computer. I. Principle of operation and model evaluation. Proceedings of the third International Conference on Medical Physics including Medical Engineering (Göteborg 1972).Google Scholar
  2. 2.
    Jones, W. B., J. B. Griffin: J. Appl. Physiol.17, 482 (1962).PubMedGoogle Scholar
  3. 3.
    Greenfield Jr., J. C., D. L. Fry: J. Appl. Physiol.20, 1141 (1965).Google Scholar

Copyright information

© Dr. Dietrich Steinkopff Verlag 1977

Authors and Affiliations

  • K. H. Wesseling
    • 1
  • P. D. Verdouw
    • 2
  • J. Beaune
    • 2
  • J. Roelandt
    • 2
  • P. G. Hugenholtz
    • 1
  1. 1.Institut of Medical Physic TNOUtrecht(The Netherlands)
  2. 2.Laboratory for Experimental Cardiology, ThoraxcenterErasmus UniversityRotterdamThe Netherlands

Personalised recommendations