Single-shot t1-and t2-weighted magnetic resonance imaging of the heart with black blood: preliminary experience

  • Michael K. Stehling
  • Nicolaus G. Holzknecht
  • Gerhard Laub
  • Dieter Böhm
  • Alexander von Smekal
  • Maximilian Reiser
Papers

Abstract

Purpose: To implement and evaluate two robust methods for T1-and T2-weighted snapshot imaging of the heart with data acquisition within a single heart beat and suppression of blood signal.

Methods: Both Tl-and T2-weighted diastolic images of the heart can be obtained with half Fourier single-shot turbo spin echo (HASTE) and turbo fast low-angle shot (turboFLASH) sequences, respectively, in less than 350 ms. Signal from flowing blood in the ventricles and large vessels can be suppressed by a preceding inversion recovery preparing pulse pair (PRESTO). Fifteen volunteers and five patients have been evaluated quantitatively for signal-to-noise ratio (SNR) contrast-to-noise ratio (CNR) and flow void and qualitatively for image quality, artifacts, and black-blood effect.

Results: Both PRESTO-HASTE and PRESTO-turboFLASH achieved consistently good image quality and blood signal suppression. In contrast to gradient-echo (GRE) echo-planar imaging techniques, (EPI) HASTE and turboFLASH are much less sensitive to local susceptibility differences in the thorax, resulting in a more robust imaging technique without the need for time-consuming system tuning. Compared to standard spin-echo sequences with cardiac triggering, HASTE and turboFLASH have significantly shorter image acquisition times and are not vulnerable to respiratory motion artifacts.

Conclusion: PRESTO-HASTE and PRESTO-turboFLASH constitute suitable methods for fast and high-quality cardiac magnetic resonance imaging (MRI).

Keywords

magnetic resonance heart rapid imaging fast spin-echo 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Bailes DR, Gilderdale DJ, Bydder GM,et al. (1985) Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artifacts in MR imaging. J Comput Assist Tomogr9: 835–838.PubMedGoogle Scholar
  2. 2.
    Hoult DI (1977) Zeugmatography: the criticism of the concept of a selective pulse in the presence of a field gradient.J Magn Reson 26: 165.Google Scholar
  3. 3.
    Stehling MK, Howseman AM, Ordidge RJ,et al. (1989) Whole-body echo-planar MR imaging at 0.5 T.Radiology 170: 257–263.PubMedGoogle Scholar
  4. 4.
    Stehling MK, Turner R, Mansfield P (1991) Echo-planar imaging: magnetic resonance imaging in a fraction of a second.Science 254: 43–50.PubMedGoogle Scholar
  5. 5.
    Rzedzian RR, Pykett IL (1987) Instant images of the human heart using a new, whole-body MR imaging system.AJR 149: 245–250.PubMedGoogle Scholar
  6. 6.
    Davis CP, McKinnon GC, Debatin JF,et al. (1994) Normal heart: evaluation with echo-planar MR imaging.Radiology 191: 691–696.PubMedGoogle Scholar
  7. 7.
    Ordidge RJ, Howseman A, Coxon R,et al. (1989) Snapshot imaging at 0.5 T using echo-planar techniques.Magn Reson Med 10: 227–240.PubMedGoogle Scholar
  8. 8.
    Schmitt F, Stehling MK, Ladebeck R,et al. (1992) Echo-planar imaging of the central nervous system at 1.0 T.J Magn Reson Imaging 2: 473–478.PubMedGoogle Scholar
  9. 9.
    Mansfield P, Pykett IL (1978) Biological and medical imaging by NMR.J Magn Reson 29: 355–373.Google Scholar
  10. 10.
    Listerud J, Einstein S, Outwater E, Kressel HY (1992) First principles of fast spin echo.Magn Reson Q 8: 199–244.PubMedGoogle Scholar
  11. 11.
    Simonetti OP, Laub G, Finn JP (1994)Proceedings of the Society of Magnetic Resonance: 1499.Google Scholar
  12. 12.
    Laub G, Simonetti O, Nitz W (1995) Single-shot imaging of the heart with HASTE.Proceedings of the Society of Magnetic Resonance Nice, France: 246.Google Scholar
  13. 13.
    Edelman RR, Chien D, Kim D (1991) Fast selective black blood MR imaging.Radiology 181: 655–660.PubMedGoogle Scholar
  14. 14.
    Roemer PB, Edelstein WA, Hayes CE,et al. (1990) The NMR phased array.Magn Reson Med 16: 192–225.PubMedGoogle Scholar
  15. 15.
    Haase A, Matthaei W, Bartkowski R,et al. (1989) Inversion recovery snapshot-FLASH MR imaging.J Comput Assist Tomogr 13: 1036.PubMedGoogle Scholar

Copyright information

© Chapman & Hall 1996

Authors and Affiliations

  • Michael K. Stehling
    • 1
  • Nicolaus G. Holzknecht
    • 1
  • Gerhard Laub
    • 3
  • Dieter Böhm
    • 2
  • Alexander von Smekal
    • 1
  • Maximilian Reiser
    • 1
  1. 1.Department of Diagnostic RadiologyKlinikum Großhadern der Ludwig-Maximilian-Universität MünchenMünchenGermany
  2. 2.Department of Cardiac SurgeryLudwig-Maximilian-UniversityMunichGermany
  3. 3.Siemens Medical Engineering GroupErlangenGermany

Personalised recommendations