Evaluation of a simple device for regulating intravenous infusions

  • Shyam V. S. Rithalia
  • Adrian Rozkovec
Methods and Devices

Abstract

Dial-A-Flo, and infusion control device, is assessed for accuracy, in vitro and in vivo. The unit did not permit accurate control and particularly so at lower flow rates. The normal spontaneous variations in venous pressure that occur around an infusions cannula tip, further increase the error of gravity-fed infusions. Dial-A-Flo has no advantage over a standard administration set alone, and should not be used for controlled administration of cardiac drugs.

Key words

Intravenous Infusion Device 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    LaCour, D.: Drop size in intravenous infusion. Acta Anaesthesiol. Scand.24, 35 (1966)Google Scholar
  2. 2.
    Ferenchak, P., Collins Jr., J.J., Morgan, A.: Drop size and rate in parenteral infusion. Surgery70, 674 (1971)Google Scholar
  3. 3.
    Flack, F.C., Whyte, T.D.: Behavour of standard gravity-fed administration sets used for intravenous infusion. Br. Med. J.3, 439 (1974)Google Scholar
  4. 4.
    Demoruelle, J.L., Harrison, W.L., Flora, R.E.: Flow rate maintenance and output of intravenous fluid administration sets. Am. J. Hosp. Pharm.32, 177 (1975)Google Scholar
  5. 5.
    Rithalia, S.V.S., Tinker, J.: Automatic infusion devices. Br. J. Clin. Equip.3, 163 (1978)Google Scholar

Copyright information

© Springer-Verlag 1979

Authors and Affiliations

  • Shyam V. S. Rithalia
    • 1
  • Adrian Rozkovec
    • 1
  1. 1.The Intensive Therapy UnitThe Middlesex HospitalLondonEngland

Personalised recommendations