Abstract
We compared values of invasive blood pressure measured intra-arterially with those measured noninvasively with an automated oscillometric monitor. Twenty-eight patients undergoing cardiac surgical procedures under general anesthesia were studied and 552 determinations were made. The two methods of measuring blood pressure correlated within the expected bounds of experimental accuracy and physiological variation. However, the correlation between invasive and noninvasive methods varied, apparently arbitrarily, with time. These disparities could not be explained by a linear combination of physiological variables recorded. Systolic determinations correlated the best and diastolic the least between the invasive and noninvasive methods. In general, the correlation was better for adults than for children, except with diastolic blood pressure.
Similar content being viewed by others
References
AAMI Draft Standard for Electronic or Automated Sphygmomanometers. Association for the Advancement of Medical Instrumentation, Arlington, VA, February 1980, pp 43–44.
Friesen RH, Lichtor JL: Indirect measurement of blood pressure in neonates and infants utilizing an automated non-invasive oscillometric monitor. Anesth Analg 60: 742–745, 1981.
Geddes LA, Chaffee V, Whistler SJ, Bourland JB, Tacker WA: Indirect mean blood pressure in the anesthetized pony. Am J Vet Res 38: 2055–2057, 1977.
Geddes LA, Combs W, Denton W, Whistler SJ, Bourland JB: Indirect mean arterial pressure in the anesthetized dog. Am J Physiol 238: H664, 1980.
Geddes LA, Moore AG, Garner H: The indirect measurement of mean blood pressure in the horse. Southwest Veterin 23: 289–298, 1970.
Gravenstein JS, Newbower RS, Ream AK, Smith NT: Essential Non-invasive Monitoring in Anesthesia. Grune and Stratton, New York, 1980.
Kimble KJ, Darnell RA, Yelderman M, Ariagno RL, Ream AK: An automated oscillometric technique for estimating mean arterial pressure in critically ill newborns. Anesthesiology 54: 423–425, 1981.
Mauck GW, Smith CR, Geddes LA, Bourland JB: The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. Part II. J Biomechan Eng 102: 28–33, 1980.
Posey JA, Geddes LA, Williams H, Moore AG: The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. Part I. Cardiovasc Res Ctr Bull 8(1): 15–25, 1969.
Ramsay M: Non-invasive automatic determination of mean arterial pressure. Med Bio Eng Comput 17: 11–18, 1979.
Yelderman M, Ream AK: Indirect measurement of mean blood pressure in the anesthetized patient. Anesthesiology 50: 253–256, 1979.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Green, M., Paulus, D.A., Roan, V.P. et al. Comparison between oscillometric and invasive blood pressure monitoring during cardiac surgery. J Clin Monit Comput 1, 21–26 (1984). https://doi.org/10.1007/BF01726658
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01726658