Evaluation of an apparatus for continuous monitoring of gas exchange in mechanically ventilated patients

Abstract

Indirect calorimetry has been integrated with a commercially available ventilator, permitting continuous monitoring of energy expenditure and respiratory quotient. This equipment, the ventilator Erica, the Engström metabolic computer EMC and the carbon dioxide analyzer Eliza (Gambro Engström AB, Sweden) was evaluated in laboratory tests and in critically ill patients. In the laboratory evaluation, the variability in FIO2 was less than 0.05% at FIO2<0.47. The linearity error of the oxygen sensor and the CO2 analyzer was less than 0.06%. The variability of the oxygen sensor was 0.01% O2 and of the CO2 analyzer 0.04% CO2. The CO2 analyzer was sensitive to oxygen concentration and underestimated CO2 concentration by 3.3% at FIO2=0.5 if calibrated on room air. The pneumotachometer was oxygen dependent and overestimated inspired volume by 5.5% at 100% O2, if calibrated on room air, and was negligibly affected by changes in minute volume or PEEP within the physiological range for adults. In the patient evaluation, 15 measurements of gas exchange were performed in 9 critically ill patients and the results were compared with data obtained simultaneously by the Scholander technique or mass spectrometry. The standard deviation of the mean difference in comparison with the reference methods was 9.4% for O2 uptake, 6.6% for CO2 elimination, 12.5% for respiratory quotient and 3.9% for minute volume. Paired t-test analysis showed no systematic difference between the various methods.

It is concluded that the EMC has an accuracy sufficient for clinical use and represents a commercially available technique of potential clinical value.

References

  1. Quebbeman EJ, Ausman RK, Schneider TC: A re-evaluation of energy expenditure during parenteral nutrition. Ann Surg 195: 282–286, 1982.

    PubMed  Article  CAS  Google Scholar 

  2. Quebbeman EJ, Ausman RK: Estimating energy requirements in patients receiving parenteral nutrition. Arch Surg 117: 1281–1284, 1982.

    PubMed  CAS  Google Scholar 

  3. Askanazi J, Nordenström J, Rosenbaum SH, Elwyn DH, Hyman AI, Carpentier YA, Kinney JM: Nutrition for the patient with respiratory failure: glucose versus fat. Anesthesiology 54: 373–377, 1981.

    PubMed  Article  CAS  Google Scholar 

  4. Sheldon GF, Petersen SR, Sanders R: Hepatic dysfunction during hyperalimentation. Arch Surg 113: 504–508, 1978.

    PubMed  CAS  Google Scholar 

  5. Neuhof H, Wolf H: Method for continuously measured oxygen consumption and cardiac output for use in critically ill patients. Crit Care Med 6: 155–161, 1978.

    PubMed  Article  CAS  Google Scholar 

  6. Damask MC, Askanazi J, Weissman C, Elwyn DH, Kinney JM: Artifacts in measurement of resting energy expenditure. Crit Care Med 11: 750–752, 1983.

    PubMed  Article  CAS  Google Scholar 

  7. Carlsson M, Nordenström J, Hedenstierna G: Clinical implications of continuous measurement of energy expenditure in mechanically ventilated patients. Clin Nutr 3: 103–110, 1984.

    PubMed  CAS  Google Scholar 

  8. Biebuyck J: Total parenteral nutrition in the perioperative period — a time for caution? Anesthesiology 54: 360–362, 1981.

    PubMed  Article  CAS  Google Scholar 

  9. Gedeon A, Hamilton K, Häggmark S, Reiz S: New sampling technique for respiratory gas measurements (abstract). Intensive Care Medicine 3: 224, 1983.

    Google Scholar 

  10. Lusk G: Analysis of the oxidation of mixtures of carbohydrate and fat. J Biol Chem 59: 41–42, 1924.

    CAS  Google Scholar 

  11. Snedecor GW, Cochran WG: Statistical methods, 6th Ed. Ames, Iowa, pp 346–349, 1967.

    Google Scholar 

  12. Grenvik Å, Hedstrand U, Sjögren H: Problems in pneumotachography. Acta Anaesth Scand 10: 147–155, 1966.

    PubMed  CAS  Article  Google Scholar 

  13. Browning JA, Linberg SE, Turney SZ, Chodoff P: The effects of a fluctuating FIO2 on metabolic measurements in mechanically ventilated patients. Crit Care Med 10: 82–85, 1982.

    PubMed  Article  CAS  Google Scholar 

Download references

Author information

Affiliations

Authors

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Carlsson, M., Forsberg, E., Thörne, A. et al. Evaluation of an apparatus for continuous monitoring of gas exchange in mechanically ventilated patients. J Clin Monit Comput 1, 211 (1985). https://doi.org/10.1007/BF01720185

Download citation

  • DOI: https://doi.org/10.1007/BF01720185

Keywords

  • Oxygen uptake
  • indirect calorimetry
  • ventilator
  • patients
  • evaluation