Abstract
Objective
The measure of the precision and accuracy of three optic SvO2 systems in vitro (SAT-2 Baxter, oximetrix 3 Abbott, HEMOPRO-2 Spectramed in a wide range of saturations (24–85%) and hematocrit (20–40%).
Design
In vitro comparison of the results given by a Co-Oximeter and the SvO2 systems.
Measurements and results
144 blood samples were tonometered with a variety of gas mixtures. Measurements of SvO2 were compared to those obtained by means of a Co-Oximeter, the difference (dSO2) between paired values were calculated. Precision was evaluated from confidence intervals (2 SD) and accuracy was tested using dSO2 figures. As there was no difference in either precision or accuracy among catheters for a given system, data obtained from a given system were pooled. Mean precision was 14.2%, 8.6%, 9% for the Spectramed, Baxter and Abbott systems, respectively.
Conclusion
The rather low reliability of the optic method should be borne in mind in patients with low SvO2 and/or low hematocrit.
Similar content being viewed by others
References
Birman H, Haq A, Hew E, Aberman A (1984) Continuous monitoring of mixed venous oxygen saturation in hemodynamically unstable patients. Chest 86:753–756
Gettinger A, De Traglia MC, Glass DD (1987) In vivo comparison of two mixed venous saturation catheters. Anesthesiology 66: 373–375
Scuderi PE, Bowton DL, Meredith JW, Harris LC, Brocksch J (1992) Clinical evaluation of three mixed-venous oxygen saturation catheters. Chest 102:896–905
Rouby JJ, Poete P, Bodin L, Bourgeois JL, Arthaud M, Viars P (1990) Three mixed venous saturation catheters in patients with circulatory shock and respiratory failure. Chest 98:954–958
Reinhart K, Moser N, Rudolph T, Bredle D, Specht M, Gramm HJ, Goecke J, Eyrich K (1988) Accuracy of two mixed venous saturation catheters during long-term use in critically ill patients. Anesthesiology 89:769–773
Karis JH, Lumb PD (1988) Clinical evaluation of the Edwards Laboratories and Oximetrix mixed venous oxygen saturation catheters. Cardiothorac Anesth 2:440–444
Hecker BR, Brown DL, Wilson D (1989) A comparison of two pulmonary artery venous oxygen saturation catheters during the changing conditions of cardiac surgery. J Cardiothorac Anesth 3:269–275
Severinghaus JW (1958) Oxyhemoglobin dissociation curve correction for temperature and pH variation in human blood. J Appl Physiol 12:485–486
Bland JM, Altman DG (1985) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310
Kelman GR, Nunn JF (1966) Nomograms for correction of blood PO2, PCO2, pH, and bases excess for time and temperature. J Appl Physiol 21:1484–1490
Stetz GW, Miller RG, Kelly GE, Raffin TA (1982) Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 126:1001–1004
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Janvier, G., Guenard, H. & Lomenech, A.M. In vitro accuracy of three blood O2 saturation optic catheter systems. Intensive Care Med 20, 480–483 (1994). https://doi.org/10.1007/BF01711899
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01711899