Noise in the ICU

Abstract

Objective

The growing number of technical devices in ICUs makes noise exposure a major stressor. The purpose of this study was to assess noise levels during routine operation in our ICU.

Design

Our ICU is an open ward with four rooms, constructed in the 1960s. During the study period, 4 patients were in the controlled room and were treated by 4 nurses during the day and by 2 at night. A-weighted sound pressure levels (SPL) were measured continuously for 2 days and nights. Also measured were the alarms of various appliances. For gross overall evaluation it is customary to state the Leq, i.e. the energy-averaged level during measurement. The annoyance caused by noise depends more on rare events of high intensity. Therefore, the distribution of SPL values (Ln) over time was also analysed.

Results

SPL was roughly the same during the day and at night, with Leq between 60–65 dB(A) and peaks up to 96 dB(A). Most alarms reach an SPL of 60–70 dB(A), but some exceed 80 dB(A). During teaching rounds Leq exceeds 65 dB(A).

Conclusion

During the day and at night SPL always surpasses the permissible noise exposure for 24 h of 45 db(A) recommended by the US Environmental Protection Agency. Alarms cause the most irritating noise. Hospital management should pay attention to internal noise, and SPL should be measured routinely.

References

  1. Aitken RJ (1982) Quantitative noise analysis in a modern hospital. Arch Environ Health 37:361–364

    PubMed  Google Scholar 

  2. Sharkey T (1988) Sounding an alarm on alarms (letter). Am J Nurs 88:33

    Google Scholar 

  3. Hansell HN (1984) The behavioral effects of noise on man: the patient with “intensive care unit psychosis”. Heart Lung 13:59–63

    PubMed  Google Scholar 

  4. Falk SA, Woods NF (1973) Hospital noise levels and potential health hazards. N Engl J Med 289:774–781

    PubMed  Google Scholar 

  5. US Environmental Protection Agency (1974) Information on levels of environmental noise requisite to protect health and welfare with an adequate margin of safety. EPA, Washington DC

    Google Scholar 

  6. Field T (1990) Alleviating stress in newborn infants in the intensive care unit. Clin Perinatal 17:1–5

    Google Scholar 

  7. Thomas KA (1989) How the NICU environment sounds to a preborn infant. MCN 14:249–252

    Google Scholar 

  8. Hodge B, Thompson JF (1990) Noise pollution in the operating theatre. Lancet 335:891–894

    PubMed  Google Scholar 

  9. Hilton A (1987) The hospital racket: how noisy is your unit? Ann Y Nurs 1:59–60

    Google Scholar 

  10. Kerr JH, Hayes B (1983) An “alarming” situation in the intensive therapy unit. Intensive Care Med 9:103–104

    PubMed  Google Scholar 

  11. Shapiro RA, Berland T (1972) Noise in the operating room. N Engl J Med 287:1236–1238

    PubMed  Google Scholar 

  12. Updike P (1990) Music therapy results for ICU patients. Dimens Crit Care Nurs 9:39–42

    PubMed  Google Scholar 

  13. Pownall M (1987) Medical alarms: ringing the changes. Nurs Times 83:18–20

    Google Scholar 

Download references

Author information

Affiliations

Authors

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Balogh, D., Kittinger, E., Benzer, A. et al. Noise in the ICU. Intensive Care Med 19, 343–346 (1993). https://doi.org/10.1007/BF01694709

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01694709

Key words

  • Noise
  • Alarms
  • ICU