Cardiac output was measured by the thermodilution method in a patient with a left-to-right shunt undergoing cardiac catheterization. It appeared that the thermodilution method measured systemic rather than pulmonary blood flow. This occurred because of the slow injection of injectate in the presence of a large left-to-right shunt. Theoretical thermodilution cardiac output curves are provided to illustrate the interaction of these two factors when three different durations of injection and four different shunt sizes are used.
This is a preview of subscription content, access via your institution.
References
Carter SA, Bajec DF, Yannicelli E, Wood EH. Estimation of left-to-right shunt from arterial dilution curves. J Lab Clin Med 1960;55:77–88
Hills LD, Winniford MD, Jackson JA, Firth BG. Measurement of left-to-right intracardiac shunting in adults: oximetric versus indicator dilution techniques. Cathet Cardiovasc Diagn 1985;11:467–472
Dalen JE, Grossman W. Shunt detection and measurement. In Grossman W, ed. Cardiac catheterization and angiography. Philadelphia: Lea & Febiger, 1980:131–143
Askenazi J, Ahnberg DS, Korngold E, et al. Quantitative radionuclide cardiography: detection and quantitation of left to right shunts. Am J Cardiol 1976;37:382–387
Meyer RA. Echocardiography in pediatric patients. Cardiovasc Clin 1981;11(2):187–225
Morady F, Brundage BH, Gelberg HJ. Rapid method for determination of shunt ratio using a thermodilution technique. Am Heart J 1983;106:369–373
Alpert BS, Eubig C. Thermodilution Qp/Qs: an indicator dilution method. Pediatr Cardiol 1983;4:13–17
Billings RG, Clark JS, Veasy LG, Cutler CA. Shunt quantification by mathematical analysis of indicator dilution curves. Cathet Cardiovasc Diagn 1978;4:143–162
McGuirk SM, Welker FH, Muir WW, Hull BL. Thermodilution curves for diagnosis of ventricular septal defect in cattle. J Am Vet Med Assoc 1984;184:1141–1145
Hedvall G. The applicability of the thermodilution method for determination of pulmonary blood flow and pulmonary vascular resistance in infants and children with ventricular septal defects. Scand J Clin Lab Invest 1978;38:581–585
Jögi P, Werner O. Left-to-right shunt assessed by thermodilution during surgery for congenital heart disease. Scand J Thorac Cardiovasc Surg 1987;21:203–206
Cutler GA, Billings RG, Clark JS, Veasy LG. A thermodilution method for quantification of bidirectional shunts. Comput Biomed Res 1979;12:379–410
McMurray RG, Katz VL, Berry MJ, Cefalo RC. The effect of pregnancy on metabolic responses during rest, immersion, and aerobic exercise in the water. Am J Obstet Gynecol 1988;158:481–486
Runciman WB, Ilsley AH, Roberts JG. An evaluation of thermodilution cardiac output measurement using the Swan-Ganz catheter. Anaesth Intensive Care 1981;9:208–220
Levett JM, Replogle RL. Thermodilution cardiac output: a critical analysis and review of the literature. J Surg Res 1979;27:392–404
American Edwards Laboratories. Model 9520 and 9520A cardiac output computer operations and troubleshooting manual. Santa Ana, CA: American Edwards, 1983
Low LR, Windham JP, Farison JB, Potvin WJ. Computer model for simulation of first transit cardiac radionuclide curves—I. IEEE Trans Biomed Eng 1989;36:935–945
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Pearl, R.G., Siegel, L.C. Thermodilution cardiac output measurement with a large left-to-right shunt. J Clin Monitor Comput 7, 146–153 (1991). https://doi.org/10.1007/BF01618114
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01618114
Key words
- Measurement techniques: thermodilution
- Heart: cardiac output