Summary

Cylindrical (radius 2.5 mm) and semicylindrical (radius 6.5 mm) implants of nondegradable alumina of defined porosity (pore diameters 0.4 to 0.6 mm, 0.6 to 0.8 mm, 0.8 to 1.0 mm and 1.0 to 1.2 mm) were implanted in the tibiae of dogs. The pores were filled to an increasing extent with bone tissue, so that after a period of seven months almost all the pores contained bone tissue. Cylindrical implants (radius 2.<5 mm) of calcium phosphate with 0.8 to 1.0 mm pore diameter were more rapidly filled with bone tissue within 2 months than implants of A12O3 and of calcium phosphate-calcium alumina with 0.8 to 1.0 mm pore diameter.

Direct contact was established between the calcium phosphate implants and new bone, i.e. without any seam of connective tissue. Between the bone and the Al2O3 ceramic on the other band, there was a narrow seam of connective tissue, and a relatively broad seam was observed between the bone and the calcium phosphate-calcium alumina implant.

The calcium phosphate implants had an excellent tissue compatibility. In the vicinity of the Al2O3 implants minor, and in the case of the calcium phosphate-calcium alumina implants major foreign body reactions occurred after 2 months.

Zusammenfassung

Definiert poröse zylindrische (Radius 2,5 mm) und halbzylindrische (Radius 6,5 mm) Implantate aus resistenter Aluminiumoxidkeramik mit den Porendurchmessern 0,4 bis 0,6 mm, 0,6 bis 0,8 mm, 0,8 bis 1,0 mm und 1,0 bis 1,2 mm wurden in die Tibia von Hunden implantiert. In den Poren bildete sich in zunehmendem Maße Knochengewebe, und nach einer Verweilzeit von 7 Monaten enthielten fast alle Poren vorwiegend Knochen-gewebe. In zylindrischen Implantaten aus Calciumphosphat (Radius 2,5 mm) mit 0,8 bis 1,0 mm Porendurchmesser erfolgte die Knochenbildung innerhalb der zweimonatigen Verweilzeit deutlich schneller als in vergleichbaren Implantaten aus A1203 und in Implantaten aus Calciumphosphat-Calciumaluminat mit 0,8 bis 1,0 mm Porendurchmesser.

Zwischen Knochen und Calciumphosphat-Implantat entstand eine direkte Verbindung ohne erkennbaren Bindegewebssaum. Zwischen Knochen und Al2O3-Keramik hingegen war ein schmaler, zwischen Knochen und Calciumphosphat-Calciumaluminat ein vergleichsweise breiter Bindegewebssaum vorhanden.

Die Gewebeverträglichkeit war bei Calciumphosphat optimal. In der Umgebung von Al2O3 traten nach mehrwöchiger Verweilzeit geringgradige Fremdkörperreaktionen auf, in der Umgebung von Calciumphosphat-Calciumaluminat nach 2 Monaten starke Fremdkörperreaktionen.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  • Andrian-Werburg, H. v., Griss, P., Krempien, B., Heimke, C.: Klinische Problematik und experimentelle morphologische Befunde bei der Verwendung keramischer Werkstoffe in der Orthopädie und Unfallchirurgie. Z. Orthop.111, 577–579 (1973)

    Google Scholar 

  • Hulbert, D. F., Klawitter, J. J., Sauer, B. W., Matthews, J. R.: Characterization of tissue ingrowth into porous bioceramics. Technical Report No. 2 for Office of Naval Research. Clemson: Clemson University 1972

    Google Scholar 

  • Ruckelshausen, M. C. v.: Haltbarkeitserwartung von Hilfttotalprothesen in Abhängigkeit von Verschleiß, Reibungskoeffizient und Kopfdurchmesser. In: Der totale Hiiftgelenkersatz (Hrsg. H. Cotta, K. P. Schulitz). Stuttgart: Thieme 1973

    Google Scholar 

  • Schuster, J.: Metallurgische Probleme in der Knochenchirurgie. Munch. med. Wschr.112, 1590–1594 (1970)

    Google Scholar 

  • Talbert, C. D., Klawitter, J. J., Hulbert, S. F.: Materials of construction for bone gap bridges. Clemson: Clemson University 1969

    Google Scholar 

  • Willert, H. G., Semlitsch, M.: Gewebsreaktionen um derzeit gebräuchliche Hilftendoprothesen. In: Der totale Hiiftgelenkersatz (Hrsg. H. Cotta, K. P. Schulitz). Stuttgart: Thieme 1973

    Google Scholar 

  • Willert, H. G.: Die Reaktion des knöchernen Implantatlagers auf Methylmethacrylatknochenzement. In: Der totale Hüftgelenkersatz (Hrsg. H. Cotta, K. P. Schulitz). Stuttgart: Thieme 1973

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Die Untersuchungen sind Teil einer Forschungsarbeit, die das Battelle-Institut, Frankfurt a. M., im Auftrag des Bundesministeriums für Forschung und Technologie durchführt.