Summary
From general physical laws simple formulae have been deduced from which the rigidity of an external fixation unit mounted in different ways becomes mathematically calculable. Furthermore, the stability of various mounting systems under bending forces was investigated experimentally.
Particular attention was paid to details of operating technique which are of special importance to the stability of the whole construction. The calculated data are consistent with the mechanical test results.
The best arrangement of the Steinmann pins and Schanz screws is discussed, and how to mount the frame optimally, depending on the specific type and location of the fracture. Special instructions concerning the mounting of the external fixation unit are given which, depending on the specific circumstances, guarantee optimum stability of the unit.
Zusammenfassung
In vorliegender Arbeit wurden aus allgemein gültigen physikalischen Gesetzmäßigkeiten einfache Formeln abgeleitet, mit deren Hilfe die Festigkeit verschiedener Montageformen eines äußeren Spanners rein rechnerisch ermittelt werden kann. Zusätzlich wurden these Anordnungen im Experiment auf ihre Stabilitat gegen Biegekräfte geprüft. Besondere Aufmerksamkeit wurde jenen operationstechnischen Details gewidmet, deren Beachtung für die Festigkeit der gesamten Konstruktion von Bedeutung ist. Die berechneten Werte zeigten eine gute Übereinstimmung mit den experimentell ermittelten Ergebnissen.
Auf these Weise konnten für verschiedene Fraktur-lokalisationen und -formen jeweils ideale Anordnungen der Steinmann-Nägel and Schanzschen Schrauben sowie optimale Montageformen des Rahmens gefunden und Richtlinien ausgearbeitet werden, deren Beachtung bei einwandfreier Operationstechnik eine unter den jeweiligen Bedingungen optimale Stabilität des Spanners gewährleistet.
Similar content being viewed by others
References
Burger, H., Kraus, J., Hild, P., Hofmann, D.: Festigkeitsuntersuchungen am Fixateur externe unter Biegebeanspruchung bei Defekten am Bruchspalt. Unfallchirurgie3, 221–225 (1977)
Diehl, K.: Stabilität und Biegebeanspruchung von Osteosynthesen des Ober- und Unterschenkels bei der Frühmobilisation. Unfallheilkunde79, 81–89 (1976)
Hierholzer, G., Kleining, R., Hörster, G.: Osteosynthese mit dem Fixateur externe. Unfallchirurgie3, 209–219 (1977)
Hierholzer, G., Kleining, R., Hörster, G., Zemenides, P.: External fixation. Classification and indications. Arch. Orthop. Traumat. Surg. 92, 175–182 (1978)
Hild, P., Hofmann, D., Burger, H., Kraus, J.: Festigkeitsuntersuchungen am Fixateur externe unter Biegebeanspruchung. Unfallchirurgie4, 77–83 (1978)
Hofmann, D., Burger, H., Kraus, J., Hild, P.: Festigkeitsuntersuchungen am Fixateur externe unter Biegebeanspruchung bei Defekten am Bruchspalt. Unfallchirurgie3, 147–153 (1977)
Kleining, R., Hierholzer, R.: Biomechanische Untersuchungen zur Osteosynthese mit dem Fixateur externe. Akt. Traumat.6, 71–76 (1976)
Kraus, J., Hild, P., Hofmann, D., Burger, H.: Festigkeitsuntersuchungen am Fixateur externe unter Biegebeanspruchung. Unfallchirurgie3, 227–231 (1977)
Labitzke, R., Henze, G.: Biomechanik des Fixateur externe. Unfallheilkunde81, 546–552 (1978)
Sass, F., Bouche, Ch.: Dubbel's Taschenbuch fur den Maschinenbau, 13. Aufl. Berlin, Heidelberg, New York: Springer 1974
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Egkher, E., Martinek, H. & Wielke, B. How to increase the stability of external fixation units. Arch. Orth. Traum. Surg. 96, 35–43 (1980). https://doi.org/10.1007/BF01246140
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01246140