Improved sensitivity of the iodine pentoxide method for the estimation of carbon monoxide

The use of silica gel as an adsorbent for iodine vapour

  • Published:
Microchimica Acta Aims and scope Submit manuscript

Summary

A simple method of improving the sensitivity of the iodometric estimation of carbon monoxide is described. Iodine vapour is trapped on dry silica gel granules instead of in potassium iodide solution as in the conventional methods. Titration of the adsorbed iodine can be conducted in a medium of optimum iodide and starch concentration, the colour change due to adding one microlitre aliquots of N/560 sodium thiosulphate, being detectable. The new technique is particularly suitable for the estimation of less than 10 p. p. m. of carbon monoxide in gas samples of about 300 ml volume, and is also useful as a rapid means of checking the apparatus “blank” at all times.

Zusammenfassung

Die Empfindlichkeit der jodometrischen Kohlenmonoxydbestimmung läßt sich in einfacher Weise verbessern, indem man die Joddämpfe statt wie üblich in eine Kaliumjodidlösung über trockenes Silikagel leitet. Die Titration des adsorbierten Jods kann dann bei optimaler Konzentration an Jodid und Stärke ausgeführt werden und ist infolge des auf ± 1 μl 1/560−n Thiosulfatlösung scharf erkennbaren Umschlagpunktes genau durchführbar. Das neue Verfahren ist besonders für die Bestimmung von weniger als 0,01 Vol.-% Kohlenoxyd in Gasproben von etwa 300 ml geeignet. Außerdem läßt sich damit jederzeit der Blindwert der Apparatur rasch ermitteln.

Résumé

On décrit une méthode simple pour améliorer la sensibilité du dosage iodométrique de l'oxyde de carbone. La vapeur d'iode est captée sur des grains de silicagel sec à la place de la solution d'iodure de potassium comme dans les méthodes classiques. Le titrage de l'iode adsorbé peut être conduit dans un milieu de concentration optimale en iode et amidon, le changement de coloration dû à l'addition d'aliquotes de thiosulfate de sodium N/560 étant décelable. La nouvelle technique est particulièrement convenable pour le dosage de moins de 10 p. p. m. d'oxyde de carbone dans les échantillons de volume 300 ml environ et se montre utile également comme moyen rapide de contrôle à tout moment de l'essai à blanc.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literature

  1. A. Ditte, Bull. soc. chim. France13, 318 (1870).

    Google Scholar 

  2. J. S. Haldane andJ. I. Graham, Methods of Air Analysis. London: Griffin. 1935.

    Google Scholar 

  3. J. I. Graham andF. Lawrence, Chem. and Ind.3, 165 (1925).

    Google Scholar 

  4. J. Tausz andK. Jungmann, Gas- u. Wasserfach70, 1049 (1927).

    Google Scholar 

  5. G. Ljunggren andG. Frang, Svensk Kem. Tidskr.44, 279 (1932).

    Google Scholar 

  6. E. C. Roberson, J. Soc. Chem. Ind.57, 39 (1938).

    Google Scholar 

  7. L. Shnidman, Amer. Gas Assoc. Month.25, 405, 428, 503, 522 (1943).

    Google Scholar 

  8. E. G. Adams andN.T. Simmons, J. Appl. Chem. London1951, Suppl. 1, 20.

  9. M. Shepherd, J. Res. Nat. Bur. Stand.38, 351 (1947), R. P. 1777.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Glover, H.G. Improved sensitivity of the iodine pentoxide method for the estimation of carbon monoxide. Mikrochim Acta 43, 5–10 (1955). https://doi.org/10.1007/BF01220645

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01220645

Keywords

Navigation