Angiographic assessment of dimensions of 6F and 7F Mallinckrodt Softouch® coronary contrast catheters from digital and cine arteriograms

  • Research Articles
  • Published:
The International Journal of Cardiac Imaging Aims and scope Submit manuscript

Abstract

Coronary contrast catheters are almost exclusively used for calibration purposes in quantitative coronary arteriography. In this study we have assessed the suitability of new 6F and 7F Mallinckrodt nylon catheter with Softouch® tip with improved imaging specifications for such calibration purposes both from digital and cinefilm images using new analytical QCA-software packages (Philips DCI/ACA and Medis CMS). The average signed differences between the angiographically measured dimensions at 100% contrast fillings and acquired at 3 different kV-levels (60, 75 and 90 kV) were −3.3% and 0.6% for the 6F and 7F catheter tips, respectively as measured with the ACA-package on digital images, and −0.4% and 2.1%, respectively, as measured with the CMS-system on cinefilm images. The pooled standard deviations were 0.102 mm and 0.107 mm for the 6F and 7F catheter tips, respectively, as measured with the ACA-package, and 0.080 mm and 0.083 mm, as measured with the CMS-system. The deviations for the nylon shafts were much larger. It became also clear that neither the filling of the catheters, nor the kV-level used, had any appreciable effect on the measurement accuracy for the Softouch® tips, which facilitates the frame selection in QCA-studies. From these data it can be concluded that the nontapering parts of the 6F and 7F Mallinckrodt Softouch® tips are very well suitable for QCA calibration purposes, but that the nylon shafts are not

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Abbreviations

ACA:

Automated Coronary Analysis

DCI:

Digital Cardiac Imaging

QCA:

Quantitative Coronary Arteriography

CMS:

Cardiovascular Measurement System

References

  1. Reiber JHC, Kooijman CJ, den Boer A, Serruys PW. Assessment of dimensions and image quality of contrast catheters from cineangiograms. Cath Cardiov Diagn 1985; 11: 512–31.

    Google Scholar 

  2. Reiber JHC, van der Zwet PMJ, von Land CD, et al. Quantification of coronary angiograms with the DCI system. Medicamundi 1989; 34: 89–98.

    Google Scholar 

  3. van der Zwet PMJ, Pinto IMF, Serruys PW, Reiber JHC. A new approach for the automated definition of path lines in digitized coronary angiograms. Int J Cardiac Imaging 1990; 5: 75–83.

    Google Scholar 

  4. Reiber JHC, van der Zwet PMJ, von Land CD, et al. Quantitative coronary arteriography: Equipment and technical requirements. In: Reiber JHC, Serruys PW, (eds). Quantitative coronary arteriography. Kluwer Academic Publishers, Dordrecht, 1992 (in press).

    Google Scholar 

  5. Reiber JHC, Serruys PW, Kooijman CJ, et al. Assessment of short-, medium-, and long term variations in arterial dimensions from computer-assisted quantitation of coronary cineangiograms. Circulation 1985; 71: 280–8.

    Google Scholar 

  6. Leung W-H, Sanders W, Alderman EL. Coronary artery quantitation and data management system for paired cineangiograms. Cath Cardiov Diagn 1991; 24: 121–34.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Koning, G., van der Zwet, P.M.J., von Land, C.D. et al. Angiographic assessment of dimensions of 6F and 7F Mallinckrodt Softouch® coronary contrast catheters from digital and cine arteriograms. Int J Cardiac Imag 8, 153–161 (1992). https://doi.org/10.1007/BF01137536

Download citation

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01137536

Key words

Navigation