Conclusions
From the results of measurement of the activity of the radionuclide Mn54 in the source, carried out by the coincidence and 4π-counting method, the average value determined for the product PkWk=0.201±1.6%.
If we assume that Pk=0.9±1%, then we obtain Wk=0.223±2.6%. This value is in excellent agreement with the theoretical value of Wk=0.228 in [5] and with the experimental data obtained in [4, 6], but differs from the results obtained in [2, 7, 8] where Wk=0.262–0.280. Using the measured results of the activity of the Mn54 nuclide by the 4π(e, x)-γ-coincidence method, it can be shown that Wk cannot assume values from 0.262 to 0.280.
Thus, in the case when the source was closed from both ends by a film with a thickness of ∼0.00005 g/cm2, it was found that
from the ratios of the count rate in the coincidence channel and the γ-channel, where ε0 is the efficiency of the 4π-counter to Auger electrons taking into account the absorption in the film and in the source.
From Eq. (8) we have PkWk(1−f A )≤0.189. The relative absorption in the film in this case is ∼7%, i.e., 1>(1−f A )>0.93 and, consequently, PkWk≤0.205. If we take Pk= 0.9±1%, when Wk≤0.228.
Similar content being viewed by others
Literature cited
K. Friedrich, Atomkernenergie,8, No. 11 (1963).
M. Leisther and K. Friedrich, Atomkernenergie,10, No. 7 (1965).
A. Williams and P. I. Campion, Intern. J. Appl. Rad. and Isotopes,14 (1963).
P. Kramer, Physica,28 (1962).
E. H. Burhop, Phys. Rad.,16 (1955).
R. W. Fink and R. B. Moler, Byull. Am. Phys. Soc., Vol. 6 (1961).
C. E. Roos, Phys. Rev.,105, No. 3 (1957).
L. E. Bailey and I. B. Swedlund, Phys. Rev.,158, No. 6.(1967).
Additional information
Translated from Izmeritel'naya Tekhnika, No. 11, pp. 67–68, November, 1971.