Conclusions
The above semiconductor thermometer with a special transducer is convenient for measuring the temperature of flat, spherical, and fashioned outer and inner surfaces.
The special transducer provides temperature measurements lasting not more than 10 sec. The error in measuring temperature does not exceed ±0.3°.
The above technique for determining corrections of temperature deviations from the normal of the article and the measuring instrument by means of a semiconductor thermometer and special graphs raises the accuracy of measurement and productivity in manufacturing precision articles on metal-working machines, and reduces the percentage of rejects due to inaccurate measurements.
The above technique and instrument can be used under workshop conditions at the bench, in the Technical Control Division (OTK), inspection and test points, in factory test laboratories, and in GKLs for measuring equipment of the State Committee of Standards, Measures, and Measuring Instruments of the USSR.