Abstract
We describe standard resistance thermometers made of a rhodium-iron alloy designed for operation in the temperature range 0.02-2 K. The sensitive element of the thermometer is filled with liquid helium-4 or helium3, which improves the heat exchange by several orders of magnitude. This makes it possible to significantly increase the measuring current and to use conventional technology for precise measurement of resistance.
Similar content being viewed by others
References
G. Schuster, TMCSI, 6, Pt. 1, 449 (1992).
G. Schuster, TMCSI, 6, Pt. 1, 1009 (1992).
O. V. Lounasmaa, Principles and Methods for Obtaining Temperatures Below 1 K [Russian translation], Mir, Moscow (1977).
R. V. Duncan and G. Ahlers, TMCSI,6, Pt. 1, 243 (1992).
O. Avenel et al., Phys. Rev. Lett.,31, 76 (1973).
J. H. Bishop, A. C. Mota, and J. C. Wheatley, in: Proceedings, Thirteenth International Conference of Low Temperature Physics, Plenum, New York (1973).
Additional information
Translated from Izmeritel'naya Tekhnika, No. 12, pp. 29–31, December, 1995.
This work was done with the support of the International Science Foundation (grant No. J 21100).
Rights and permissions
About this article
Cite this article
Astrov, D.N., Korostin, S.V. & Polunin, S.P. Thermometer working standard for temperatures below 0.5 K. Meas Tech 38, 1356–1359 (1995). https://doi.org/10.1007/BF00990494
Issue Date:
DOI: https://doi.org/10.1007/BF00990494