Phase-sensitive voltmeter for determining hysteresis loops by means of an alternating current

  • E. N. Chechurina
Electrical Measurements
  • 22 Downloads

Conclusions

Tests of an experimental model of the voltmeter made at the VNIIM have shown that it is suitable for measuring the mean values of voltages with a form factor varying from 1.11 to 2.5. The mean referred error of the voltmeter determined at a frequency of 1000 cps, does not exceed ±2%. Additional frequency errors in the range of 400 to 10,000 do not exceed ±2%.

Keywords

Physical Chemistry Analytical Chemistry Experimental Model Form Factor Hysteresis Loop 